beautypg.com

3 calibration – Teledyne 2750 - Portable turbine generator gas analzyer User Manual

Page 22

background image

Installation Model

2750

Teledyne

Analytical

Instruments

22

Note: Always use two stage regulators on incoming gas lines

from high-pressure sources.

Sample pressure should be set between 0.5—30 psi, typically

5

psig

at the analyzer. CAUTION: HIGH FLOW RATES CAN DAMAGE
THE SENSOR.

3.3 Calibration

The Model 2750 has been calibrated at the factory but will require

periodic recalibration (zero and span).

The calibration gases used to calibrate the instrument are:

Zero: 100%

CO

2

for 0-100% H

2

in CO

2

and 0-

100% Air in CO

2

ranges

100% Air for 80-100% H

2

in Air range

Span:

100% air for 0-100% Air in CO

2

range.

100%

H

2

for both the 0-100% H

2

in CO

2

and 80-100% H2 in Air ranges

For the best accuracy, the analyzer should be calibrated with the

same gases that will be used to service the generator. This should
compensate for most small impurities that may be present in the service
gases

To calibrate the analyzer, both zero and span values must be set or

verified for each range. Always use the calibration switch position to
verify the zero setting on each range. This is required because the
microcontroller cannot see negative offsets. Use the procedures in
Sections 3.3.1 and 3.3.2 for calibrating the Model 2750.

 Gas to be analyzed or being used for calibration needs to be

pressure regulated to approximately 5 psig and the flow kept
below 1 SFCH or the cell could be damaged

 If the Trim Pots are adjusted too far past 100%, the reading will

return to 0%. If this happens, turn the trim pot back until the
reading drops below 100% and then make the final setting.

Note: Perform a zero calibration before span calibration.