AD-4405 Weighing Indicator
18.
18.
18.
18. Relay Output & Control Input (O
Relay Output & Control Input (O
Relay Output & Control Input (O
Relay Output & Control Input (OP
P
P
P-
-
-
-05)
05)
05)
05)
Replacing the RS-232C interface with this option, 3-relay outputs and 3-control inputs
can be used with the RS-232C interface of this option.
RS-232C functions and pin connections are the same as the RS-232C interface
described in “16. RS-232C Interface”.
The control inputs can control the indicator from an external terminal just like the front
panel key operations.
Set the external control functions at f13, f14 and f15 of the F-Functions.
When connecting each function pin to the common pin, the indicator makes the action.
Keep a signal width more than 100 ms for the On-time and Off-time.
ON(Make)
More than 100 ms More than 100 ms
OFF(Break)
The relays output the result of comparison.
Solid-state-relay
Maximum voltage
DC50V
Maximum current
DC100mA
Maximum resistance 8
9
Pin connections (DIN 8pin connector)
Function
Pin No. Signal name
Direction
Description
1
HI
Output
Relay output HI
6
OK
Output
Relay output OK
4
LO
Output
Relay output LO
Relay
output
8 COM(out) -
Relay
common
terminal
3
EXT1
Input
Control input 1 (f13)
5
EXT2
Input
Control input 2 (f14)
7
EXT3
Input
Control input 3 (f15)
Control
input
2
COM(in)
-
Control input common
Adaptable connector
JA-TCP0586 (of accessory)
Circuits
Circuits
Circuits
Circuits
HI
OK
LO
COM(out)
K
EXT3
EXT2
EXT1
+7 to 9V
680
9
680
9
680
9
COM
(in)
DIN connector
of OP-05
Control input
Relay output
18. Relay Output & Control Input (OP-05)