A&D Weighing Indicator AD-4405 User Manual
Page 28

Page
26
AD-4405 Weighing Indicator
Comparator
Item
Parameter
Description
* 0
Not used
1
Upper/lower limit comparison (2 Limits setting)
2
Upper/lower limit comparison (Target and allowance)
3
Upper/lower limit comparison (Target and % allowance)
4
HH/Hi/OK/Lo/LL comparison (4 Limits setting)
5
HH/Hi/OK/Lo/LL comparison (Target and allowance)
6
HH/Hi/OK/Lo/LL comparison (Target and % allowance)
7
Check weighing 1
8
Check weighing 2
9
Check weighing 3
10
Simple batch 1
11
Simple batch 2
f22 
Comparator function 
12
Simple batch 3 (Loss in weight)
10 11 Includes zero band(1) 
20 21 Includes minus(1) 
30 31 Stable(0) /All of the time(1) 
f23 (f22 1 to 6) 
 
[Type2]
Validation of 
comparison 
40 41 All of the time(0)/Start, stop(1)
Initial settings are 
11,21,31 and 40. 
10 11 Tare when start(1) 
20 21 Stop after Full by key(0)/Auto(1) 
30 31
Over/Under judgment: 
After dribble flow(0) / 
Batch finish(1) 
f23 (f22 10 to 12) 
 
[Type2]
Sub function for 
batch weighing 
40 41 No function
Initial settings are 
11,21,31 and 40. 
* 0
Over
1
Under
f24 
Relay output 
selection 
2
Finish / Full (f22 9,10,12)
Effective when 
f22 7,9,10 or 12. 
f25 
Batch finish output time 
0.0 to
9.9
Relay on time by 0.1second step 
(00 : continuous to zero band) 
Initial setting is 0.0.
f26 
Zero band 
-999999 to 999999
Initial setting is 0.
*:
Initial
settings.
 
Hold 
Item
Parameter
Description
** 0
Not hold (Hold function is off)
1
Manual hold
2
Auto hold
f27 
Hold mode 
3
Manual and Auto hold
f28 
Averaging time 
0.0 to
9.9
Averaging time by 0.1second step 
(00 : hold at start time) 
Initial setting is 0.0.
*:
Initial
settings.