A&D Weighing Indicator AD-4405 User Manual
Page 15

AD-4405 Weighing Indicator
Page
13
Dual Range
¯
¯
¯
For the range function, refer to “5.3. Weighing Range Function”.
Specifying
Specifying
Specifying
Specifying the Range and Unit
the Range and Unit
the Range and Unit
the Range and Unit
Step 1
The range and unit of measure are displayed.
Range display :
5ingl : single range
dUal : dual range
To change the range function, use the
CLEAR
key
Unit display:
The active unit is displayed. When using units that can be
used interchangeably, such as kg or lb, calibration should be
done by using the displayed unit.
To select a unit for weighing, use the
MODE
key and to select unit for
calibration, use the +/- key. The unit for calibration (first unit) is
displayed and the alternate unit (second unit) is blinking.
ENTER
Stores the value displayed and then advances to the next step.
ESC
Advances to the next step without changing the parameter.
Specifying
Specifying
Specifying
Specifying the Resolution, Decimal Point Position and Format
the Resolution, Decimal Point Position and Format
the Resolution, Decimal Point Position and Format
the Resolution, Decimal Point Position and Format
Step 2 The resolution will be displayed as d 0.1, with decimal point. Triangle 1 and the first
unit selected at the previous step will be displayed.
Positioning of the decimal point is done with the
CLEAR
key, setting of display
format (point or comma) with the +/- key, the resolution with the
MODE
key. The
decimal point format, set at this stage, will only apply to the display. The decimal
point format for serial data output is selected using the F-function settings.
Press the
ENTER
key to store the displayed settings and then proceed to the next
step. Using the
ESC
key, regardless of what is displayed, the indicator will
proceed to the next step without changing the set data.
5.2. Calibration Procedure