Constant voltage tests – Agilent Technologies N5700 User Manual

Page 94

background image

Verification

94

Series N5700 User’s Guide

Constant Voltage Tests

NOTE

Refer to the appropriate test record form for the instrument settings of the
model you are checking.

Voltage Programming and Readback Accuracy

Test category = performance, calibration

This test verifies that the voltage programming and measurement
functions are within specifications.

1 Turn off the power supply and connect a DVM directly across the

+S and -S terminals as shown in figure A. Do not connect a load.

2 Turn on the power supply and program the output voltage to zero

and the output current to its maximum programmable value
(Imax) with the load off. The CV annunciator should be on and
the output current reading should be approximately zero.

3 Record the output voltage readings on the digital voltmeter

(DVM) and the front panel display. The readings should be within
the limits specified in the test record card for the appropriate
model under Voltage Programming and Readback @ 0 V.

4 Program the output voltage to its full-scale rating.
5 Record the output voltage readings on the DVM and the front

panel display. The readings should be within the limits specified
in the test record card for the appropriate model under Voltage
Programming and Readback @ Full Scale.

CV Load Effect

Test category = performance

This test measures the change in output voltage resulting from a
change in output current from full load to no load.

1 Turn off the power supply and connect a DVM and an electronic

load as shown in figure A.

2 Turn on the power supply and program the output current to its

maximum programmable value (Imax) and the output voltage to
its full-scale value.

3 Set the electronic load for the output’s full-scale current. The CV

annunciator on the front panel must be on. If it is not, adjust the
load so that the output current drops slightly.

4 Record the output voltage reading from the DVM.
5 Open the load and record the voltage reading from the DVM

again. The difference between the DVM readings in steps 4 and 5
is the load effect, which should not exceed the value listed in the
test record card for the appropriate model under CV Load Effect.