Control company, Traceable, Manometer/pressure/vacuum instructions – Control Company 3460 MANOMETER/PRESSURE/VACUUM User Manual
Page 2
CONTROL COMPANY
4455 Rex Road
Friendswood, TX 77546 USA
Ph. 281-482-1714 Fax 281-482-9448
E-mail [email protected]
www.control3.com
Control Company is ISO 9001 Quality-Certified by DNV and ISO 17025
accredited as a Calibration Laboratory by A2LA.
Cat. No. 3460/3461/3462/3463
Traceable® Manometer/Pressure/Vacuum Instructions
92-3460-00
Rev. 3 010505
Traceable® is a registered trademark of Control Company
©
2005
Control Company
TRACEABLE
®
MANOMETER/PRESSURE/VACUUM INSTRUCTIONS
(Continued from previous page)
Differential Mode
Differential pressure is the measurement of the difference
between two pressures.
EXAMPLE: Use differential pressure sensor to measure
gauge pressure by leaving one process connection open
at atmosphere and connecting the second sensor port to
your system.
Press the DIF button, “DIF” appears at the top of the LCD.
The display indicates the relative zero – only the amount
of pressure change will be indicated. Press DIF button
again and the meter returns to normal pressure mea-
surement mode.
Recording
1. Turn the meter on and press the HOLD button for 3
seconds. “o o o o” will appear on the display and the
pressure reading will be cleared to zero.
2. Press the REC button to begin recording a reading.
“REC” will appear in the lower left of the LCD, the
timer will begin counting up and the pressure value
will display.
3. To turn the record feature off press and hold the REC
button for 3 seconds.
NOTE: Only the REC, power and backlight buttons will
be operable in the Record mode.
Minimum/Maximum Readings
While in the recording mode the minimum and maximum
readings can be recalled.
1. Press the REC button and “MAX” will display in the
lower left of the LCD. The display will show the
maximum pressure value recorded and the time of
recording.
2. Pressing the REC button again displays the “MIN”
icon on the screen and the time and value of the
minimum pressure recording.
3. Another press of the REC button returns the display to
the current timer and pressure readings.
Error Messages Displayed
“Err.1” - Pressure value exceeds maximum range
“Err.2” - Pressure value is less than minimum range
“Err.3” - When operating in DIF mode, differential pressure value
is larger than maximum range
“Err.4” - Tubing was left connected while setting zero (try reset-
ting zero), if this does not clear the error the sensor or
meter is damaged.
“E10L” or “E2UL” (when operating RS232) - pressure source is
less or over range
ATTACHMENTS
Hose/tubing
RS-232 serial computer output
ABS plastic carrying case
ALL OPERATIONAL DIFFICULTIES
If this unit does not function properly for any reason, replace the
battery with a new high-quality battery (see “Battery Replace-
ment” section). Low battery power can occasionally cause any
number of “apparent” operational difficulties. Replacing the
battery with a new fresh battery will solve most difficulties.
BATTERY REPLACEMENT
Erratic readings, a faint display, no display, flashing display or
“BAT” appearing on the display are all indicators that the battery
must be replaced. To replace the battery remove the tubing from
the unit. Lay the unit face down on a clean, flat surface. Remove
the screw from the battery cover and slide the cover off in the
direction indicated. Replace the exhausted battery with a new 9
volt alkaline battery. Replace the battery cover and screw into
place. Replacement battery Cat. No. 1112.
ACCESSORIES
Cat. No. 8045 Data Acquisition System Accessory
Captures, displays, and stores readings on any PC computer.
Information can be imported into spreadsheets, word processors,
databases, and statistical program. Supplied complete with serial
cable and CD.
WARRANTY, SERVICE, OR RECALIBRATION
For warranty, service, or recalibration, contact: