Ph calibration – Nova-Tech pH 310 User Manual
Page 8

1.
If necessary
, press the MODE key to
select pH mode
. The pH indicator
appears in the upper right hand corner
of the display
.
See figur
e
2.
Rinse the probe thoroughly with
de-ionized water or a rinse solution.
Do
not wipe the pr
obe; this causes a build-up
of electr
ostatic char
ge on the glass surface.
3.
Dip the probe into the calibration buf
fer
.
The end of the pr
obe must be completely
immersed into the sample. Stir the pr
obe
gently to cr
eate a homogeneous sample.
4.
Press CAL/MEAS
to enter pH calibration
mode. The CAL
indicator lights. The
primary display will show the measur
ed
reading while the smaller secondary dis-
play will indicate the pH standar
d buf
fer
solution.
See figur
e
5.
W
ait for the measured pH value to
stabilize.
See figur
e
15
14
6.
Press ENTER to confirm calibration
.
The meter is now calibrated to the
curr
ent buf
fer
.
See figur
e
• If you ar
e performing multipoint
calibration, go to step 7.
• If you ar
e performing one-point
calibration, go to step 9.
7.
Rinse the probe with de-ionized water
or a rinse solution, and place it in the
next pH buf
fer
.
8.
Follow steps 5 to 8 for additional
calibration points.
9.
When calibration is complete,
press CAL/MEAS to return to pH
measurement mode
.
Notes
T
o
exit from pH Calibration mode without
confirming calibration,
DO NOT pr
ess
ENTER
in step 6. Pr
ess
CAL/MEAS
instead.
If the selected buf
fer value is not within
±1.0 pH fr
om the measur
ed pH value: the
electr
ode and buf
fer icon blink and the ERR
annunciator appears in the lower left cor-
ner of the display
.
T
o
limit the number of pH buf
fer values
available during calibration, see section
P4.2 on page 33.
pH
CAL
pH
READY
D
C
D
B
C
pH
CAL
pH
pH
CAL
pH
READY
B
pH
ATC
MEAS
C
A
A
pH calibration
NOTE:
W
e r
ecommend that you perform at least a 2-point calibration using standar
d
buf
fers that bracket (one above and one below) the expected sample range.
Y
ou can perform a 1-point calibration, but make sur
e that the buf
fer value is
close to the sample value you ar
e measuring.
4.3
C
If the READY
indicator has been activiated
(set up program P4.1—see page 32), the READY
annunciator lights when the reading is stable.
http://www.novatech-usa.com/Products/Laboratory-Equipment-Supply-Products Tel: (281) 359-8538 Toll Free:(86
6) 433-6682