P4.0: unit configuration – Nova-Tech pH 310 User Manual

Page 17

background image

From measurement mode

1.

Press Setup key

to enter Set Up mode.

2.

Press the


and


keys

to scr

oll thr

ough

subgr

oups until you view parameter P4.0.

See figur

e

3.

Press the ENTER

key

twice to select

parameter 4.2.

See figur

e

4.

Press the


and


keys

to select 2, 3, 4,

or 5 point pH calibration.

5.

Press the ENTER

key

to confirm selection

and to r

eturn to the subgr

oup menu.

Pr

ess the CAL/MEAS key to r

eturn to

measur

ement mode.

From measurement mode

1.

Press Setup key

to enter Set Up mode.

2.

Press the


and


keys

to scr

oll thr

ough

subgr

oups until you view parameter P4.0.

See figur

e

3.

Press the ENTER

key

to select

parameter 4.1.

See figur

e

4.

Press the


and


keys

to select the

configuration you r

equir

e.

•OFF switches the READY

indicator of

f.

•ON switches the READY

indicator on.

pH 310 meter only:

ON and HOLD

together switches the auto endpoint

featur

e on.

5.

Press the ENTER

key

to confirm selection

and to pr

oceed to step 4 of P

.4.2. Pr

ess the

CAL/MEAS key to r

eturn to measur

ement

mode.

Notes

Meter default is set for Ready indicator on,

and auto endpoint function of

f.

33

32

P4.0: Unit configuration

P4.1: READY

indicator and auto endpoint function

Pr

ogram P4.1 lets you select “READY

indicator on” to indicate when the r

eading is

stable*, or select “READY

indicator of

f” for faster meter r

esponse.

pH 310 meter only:

Pr

ogram P4.1 also lets you switch the

Auto endpoint function on

or of

f. Select auto endpoint on to “hold” the r

eading when it is stable for mor

e than

5 seconds. The display automatically fr

eezes, and the HOLD

indicator appears on

the left side of the display

. Pr
ess the HOLD key to r

elease the display and access

other functions. Select auto endpoint of

f to deactivate this featur

e.

SETUP

READY

SETUP

ON

8.5

A

A

B

B

SETUP

READY

SETUP

ON

C

C

D

D

P4.2: Selecting number of pH calibration points

Pr

ogram P4.2 lets you select the number of calibration points that appear in pH

calibration mode: 2, 3, 4, or 5. This lets the meter scr

oll thr

ough the calibration points

mor

e quickly if you r

egularly calibrate at less than 5 points.

*The READY

indicator appears and the r

eading holds until the measur

ed value exceeds the tolerance


0.02 pH; ±

0.8 mV <400; ±

1.2 mV >400). Then, the

READY

annunciator turns of

f.