Probe care and maintenance – Nova-Tech pH 310 User Manual

Page 21

background image

41

40

pH electrode activation (for glass bodied electrodes only)

W

ARNING:

Only qualified persons pr

oficient with the safe handling of danger

ous

chemicals should perform the pr

ocedur

e below

. Pr
ovide pr

oper containers, fume

hoods, ventilation, and waste disposal. Safety goggles and pr

otective clothing must

be worn while performing this pr

ocedur

e. If possible, r

eplace with another electr

ode

instead of performing this r

eactivation pr

ocedur

e.

1.

Dip or stir the pH electrode

in alcohol for 5 minutes.

2.

Leave the electrode in tap water

for 15 minutes.

3.

Dip and stir the electrode

in concentrate acid (such as HCl or H

2

SO

4

) for five

minutes.

4.

Repeat step 2.

5.

Dip and stir in strong base

(NaOH) for five minutes.

6.

Leave for 15 minutes

in distilled or deionized water

.

7.

Now test with standard calibration buf

fer solutions

to see if the electr

ode yields

acceptable r

esults.

Y

ou may r

epeat step 3 thr

ough 6 up to thr

ee times for better

response. If the r

esponse does not impr

ove, then your electr

ode is no longer

functioning. Replace with a new electr

ode—call your OAKT

ON Distributor for mor

e

information.

9. Probe Care and Maintenance

Since your pH electr

ode is susceptible to dirt and contamination, clean it every one

to thr

ee months depending on the extent and condition of use.

NOTE:

for specialty electr

ode car

e, consult the instr

uction manual included with

your electr

ode.

pH electrode storage

For best r

esults, always keep the pH bulb wet. Use the pr

otective electr

ode storage

bottle or r

ubber cap filled with OAKT

ON

electr

ode storage solution to stor

e your

electr

ode (see page 45 for or

dering information).

Also, you can stor

e in a pH 4 buf

fer

with 1/100 part of saturated KCl. Other pH buf

fers ar

e OK for storage, but NEVER

use distilled water for storage.

After measuring

1.

Rinse the pH electr

ode and r

efer

ence junction in de-ionized water

.

2.

Store the electrode as recommended above in “pH electrode storage,” or as

recommended by the manufacturer

.

3.

Prior to next use, rinse the liquid junction with de-ionized water and tap dry—

never wipe electrode.

NOTE:

If this does not r

estor

e electr

ode to normal r

esponse, see “Reactivating the

pH electr

ode” section below

.

pH electrode cleaning

Salt deposits:

dissolve the deposits by immersing the electr

ode in tap water for ten

to fifteen minutes. T

en thor

oughly rinse with distilled water

.

Oil/grease film:

wash electr

ode pH bulb gently in some deter

gent and water

. Rinse

electr

ode tip with distilled water or use a general purpose electr

ode cleaner (see

page 40 for or

dering information).

Clogged reference junction:

heat a diluted KCl solution to 60 to 80°C. Place the

sensing part of the electr

ode into the heated solution for about 10 minutes.

Allow the

electr

ode to cool in some unheated KCl solution.

Protein deposits:

pr

epar

e a 1% pepsin solution in 0.1 M of HCl. Set the electr

ode in

the solution for five to ten minutes. Rinse the electr

ode with distilled water

.

Reactivating the pH electrode

If stor

ed and cleaned pr

operly

, your pH electr

ode should be r

eady for immediate

use. However

, a dehydrated bulb may cause sluggish r

esponse. T

o

r

ehydrate the

bulb, immerse the electr

ode in a pH 4 buf

fer solution for 10 to 30 minutes. If this

fails, the electr

ode r

equir

es activation.

Never touch or rub glass bulb. Contact