Taking measurements – Nova-Tech pH 310 User Manual
Page 11

T
o
take readings:
1.
Rinse the probe with deionized or
distilled water before use
to r
emove any
impurities adhering to the pr
obe body
.
If the pH electr
ode has dehydrated, soak
it for 30 minutes in OAKT
ON electr
ode
storage solution or a 2M–4M KCl
solution.
2.
Press ON to switch on meter
.
The
MEAS
annunciator appears on the top
center of the LCD. The
A
TC indicator
appears in the lower right hand corner
to indicate
Automatic T
emperatur
e
Compensation (See page 19 to set
Manual T
emperatur
e Compensation).
See figur
e
3.
Dip the probe into the sample.
When dipping the pr
obe into the sample,
the sensor or the glass bulb of the electr
ode
must be completely immersed into the
sample. Stir the pr
obe gently in the sample
to cr
eate a homogenous sample.
4.
Allow time for the reading to stabilize.
Note the r
eading on the display
.
5.
T
o
toggle between pH and mV (or Rel
mV) r
eadings,
press the MODE key
.
21
20
Taking Measurements
Be sur
e to r
emove the electr
ode soaker bottle or pr
otective r
ubber cap on the
electr
ode befor
e measur
ement.
pH
ATC
MEAS
C
A
A
5.3
T
aking measurements with READY indicator selected on
If the READY
indicator has been activated, the
READY
annunciator lights when
the r
eading is stable*. Switch the READY
indicator on or of
f in Set up pr
ogram
P4.1—see page 32 for dir
ections.
*The READY
indicator appears and the r
eading holds until the measur
ed value exceeds the tolerance
(±
0.02 pH; ±
0.8 mV <400; ±
1.2 mV >400). Then, the
READY
annunciator turns of
f.
T
aking measurements with the auto endpoint feature selected on
NOTE:
this featur
e is available on model pH 310 only
.
When a r
eading is stable for mor
e than 5 seconds, the auto endpoint featur
e will
automatically “hold” the r
eading. The “hold” indicator appears on the left side of
the display
. Pr
ess the HOLD key to r
elease the r
eading. Switch the
Auto endpoint
featur
e on or of
f in Set up pr
ogram P4.1—see page 32 for dir
ections.
http://www.novatech-usa.com/Products/Laboratory-Equipment-Supply-Products Tel: (281) 359-8538 Toll Free:(86