0 specifications – Nova-Tech PW Series User Manual
Page 6
Page 6 of 48
© Adam Equipment Company 2007
2.0 SPECIFICATIONS
Model
PW 124
PW 184
PW 214
PW 254
Maximum Capacity
120g
180g
210g
250g
Readability
0.0001g
Tare Range
Full
Repeatability (S.D.)
0.00015g
0.0002g
Linearity (±)
0.0002g
Units of measure
grams, milligrams, carats, ounces, pounds, troy ounce, grains,
pennyweights, drams, taels (Hong Kong, Taiwan and Singapore),
mommes, tolas, ticals and newtons
Interface
RS-232 bi-directional
Operating
temperature
+ 10°C to 40°C
Power supply
Output voltage 15 VDC, 800 mA through
External Power Supply Module- supplied as standard
(Input Voltage 100–240 V, 50/60 Hz)
Calibration
Calibration with internal mass. Alternatively the user can chose to
calibrate using an external mass
External Calibration
mass
100 g
Display
Backlit LCD with dual digits (24 mm high) and capacity tracker
Draught shield
Supplied as standard (factory fitted)
Draught shield
dimensions
202 x 524 x 275 mm
Housing
Die cast aluminium housing with glass weighing chamber
Pan size
90 mm ø
Overall dimensions
(wxdxh)
265 x 524 x 275 mm
Net weight
12 kg
Applications
Weighing, Check weighing, Percentage weighing, Parts counting,
Animal / dynamic weighing, Net / total and Density determination