beautypg.com

0 error checking, 0 supervisor menus – Nova-Tech PW Series User Manual

Page 30

background image

Page 30 of 48

© Adam Equipment Company 2007

12.0 ERROR CHECKING

During weighing the balance is constantly checking to see if the balance is
operating within limited parameters. During weighing the errors likely to occur
are:

A/D counts below lowest allowed value

A/D counts above highest allowed value

A/D not operating

Maximum capacity exceeded

Other errors may be detected during special functions or operations. These
will be described in the section that applies.

Error messages and the reason for them are:

Concerning A/D counts

ERROR ADc UL

A/D counts below a limit

ERROR ADc OL

A/D counts above a preset limit

Concerning calibration

ERROR StAb

Calibration could not be completed
because the results were not stable

ERROR LO or ERROR HI

Calibration constant not within 20% of old
calibration constant

Concerning weighing

ERROR LO

Weight display is below zero by >4%max

ERROR LOAd

Weight is above maximum plus 90d

13.0 SUPERVISOR MENUS

Pressing the

[Setup] key while in normal weighing gives access to the

menus.

When [Setup] is pressed and passcodes are not enabled the

display will show the Supervisor menus. If passcodes are enabled,
the balance will ask for it by displaying “PASSCODE 0”

• If a wrong code is entered an “ERROR CODE” message will flash and

the balance will return to weighing mode

• If the passcode has been enabled and entered, the balance will allow

the operator to access the Supervisor’s menus by which the user can