Nova-Tech PW Series User Manual
Page 34
Page 34 of 48
© Adam Equipment Company 2007
BACKLIGHT
On
OFF
AUTO (default)
POWER DOWN
Set the time after which the unit will go
into Stand-by power settings, On=Enable,
OFF=Disable, If set to On-
the options are 1 to 9 minutes
FILTER
Set a value for the amount of filtering to be
done ranging from 1 to 10. A larger
number means more filtering and a slower
response.
STABILITY
Set a value to be used to determine
balance stability, set a value of 1, 2, 5 or
10d. A larger number corresponds to a
larger stable zone. Default is 5
AUTO ZERO
Can be set to On or OFF to enable the
auto-zero function. If set to On- select
from 1, 2 or 5d
The sub-menu is entered by pressing
[Enter]-
• Use the [Up] and [Down] keys to increase or decrease the value. Press
[Enter] to accept the setting and advance to the next item in the menu
• Press [Mode] to advance to setting of the next parameter or [Esc] to
return to normal weighing
13.5 CALIBRATION SETUP
This menu allows the user to set the calibration parameters.
• Press [Enter] when “CAL SETUP” is displayed to select the calibration
parameters
• The options for each parameter can be scrolled through by using the
[Up] or [Down] key
ENABLE
NO =operator calibration is disabled
YES=operator calibration is enabled