12.3.4

F4 –INITIAL ZERO RANGE


To enter this parameter, press the [Z/T] key when “F4 int” is shown.
The display will show the current initial zero range.
Press the [Pcs/ ] key to change the value and press [Z/T] to accept the value.
Press [Print/M+/esc] to return to weighing.

12.3.5

F5 -RE-ZERO RANGE

To enter this parameter, press the [Z/T] key when “F5 rEZ” is shown.
The display will show the current re-zero range.
Press the [Pcs/ ] key to change the value.
Press [Z/T] to accept the value.
Press [Print/M+/esc] to return to weighing.

12.3.6

F6 -SUCCESSIVE TARE

To enter this parameter, press the [Z/T] key when “F6 SCS” is shown.
The display will show if the successive tare is on or off.
Press the [Pcs/ ] key to change the value.
Press [Z/T] to accept the displayed value.
Press [Print/M+/Esc] to return to weighing.

12.3.7

F7 –ADC COUNTS

To enter this parameter, press the [Z/T] key when “F7 Cnt” is shown.
This parameter allows you to view the A/D counts from the internal A/D converter.
This can be an aid to service.
Press the [Z/T] key to return to the PARAMETER menu.
Press the [Print/M+/Esc] key to return to weighing.
Typical value at zero is 30,000-90,000 (approx.).
Typical value at full capacity is 500,000 (approx.).

12.3.8

F8 –ZERO MODE

To enter this parameter, press the [Z/T] key when “F8 ZEm” is shown.
Select the Zero mode desired. In all but special cases Zero Mode 1 is used. The
other 2 zero modes are for unique locations in the world and effect the +/- range of
the zero.
Press the [Pcs/ ] key to change the value.
Press [Z/T] to accept the displayed value.
Press [Print/M+/Esc] to return to weighing.