6.8
ANIMAL
(D
YNAMIC
)
WEIGHING
The scale can be set to animal (dynamic) weighing for weighing any items that are
unstable or may move. See Section 7.4.
The scale will use a special filter to minimise the effects of any movement while the
moving animal or unstable samples are on the scale.
• Press [Func/C]. The first option is “FunC 1”, press the [Func/C] key again to
display “Func 2”.
• Press the [Z/T] key. “F2 PCt” will be displayed. Press the [Pcs/
] key to
advance to the second function, “P2 AnL”, Animal weighing.
• Press [Z/T] to enter animal weighing. The scale will change to animal weighing
and allow weighing of an unstable animal or sample on the platform.
• To use the Animal Weighing function it is necessary to set the amount of
filtering required for the item to be weighed. More active animals will
require a higher level of filtering to give a stable result. The display will show
“Flt x” where x is a value from 1 to 5. The higher the value the greater the
amount of filter will be. To increment the value shown press the [Pcs/ ] key
then press the [Z/T] key to accept it.
• The display will flash “Ani “ 2 times then show the current weight, 0.00. The
scale is now ready to weigh.