Nova-Tech AE402 Indicator User Manual
Page 18
16 |
P a g e
© Adam Equipment Company 2010
6.7
PERCENT
WEIGHING
The scale can be set to perform percent weighing. See Section 7.2.
The scale will use a mass on the platform as the 100% reference weight. If the
platform is empty (or the scale is tared) then the user can input a reference weight
using the keypad.
• Press [Func/C]. The first option is “FunC 1”, press the [Func/C] key again
display “FunC 2”.
• Press the [Z/T] key. “P1 PCt” will be displayed.
• Press [Z/T] again to enter percent weighing. The scale will set the sample
mass on the platform as 100% reference weight.
• Remove the sample weight. Then any other weight placed on the scale will be
displayed as a percentage of the original sample. For example, if 3500g is
placed on the scale and percent weighing is selected, the display will show
100.0%. Remove the 3500g weight and place a 3000g weight. The display will
show 85.7% as 3000g is 85.7% of 3500g.