• The number of decimal points will depend on the weight used in comparison

to the capacity of the system. A smaller weight will show only “100%” while a
larger weight might show “100.00%”.

• If the scale was showing zero weight when entering this function, then the

user must manually enter the weight to be set as 100% as described below.

• Press the [Z/T] key. “P1 PCt” will be displayed.

• Press [Z/T] again to enter percent weighing. The scale will now accept a value

the user enters as the reference weight.

• Enter the value using the [Unit/

] key and [Pcs/ ] key to set the values of

the 100% mass.

• When set press the [Z/T] key to store the value. The display will show “0.00

%”.

• The weight entered must be greater than 50 scale divisions.

• Press [Unit/

] to return to weighing.


NOTE:

The display may jump by large numbers unexpectedly if small weights are used to
set as 100% reference. The scale checks if the weight is too small and will show
Error 7.