Laurel Electronics LAUREATE SERIES DPM User Manual
Page 37
18.3
BAND DEVIATION
When deviation is selected from the setup menu, a value is entered for the amount of deviation
required. This value represents the number of counts at which the relay will be energized above
and below the setpoint. For example, if the setpoint is set to 10,000 and a deviation value of
200 was entered, the relay will activate below 9800 and above 10,200.
18.4
HYSTERESIS
When hysteresis is selected from the setup menu, a value is entered for the amount of
hysteresis required. This value represents the number of counts at which the relay will be
energized above and deenergized below the setpoint. For example, if the setpoint is set to
10,000 and a hysteresis value of 200 was entered, the relay will activate at 10200 and
deactivate at 9800.
18. 3 VIEWING AND CHANGING SETPOINTS
When viewing or changing the setpoint values, it is not necessary to enter the setup menu. This
allows the meter to continue conversions and provide outputs when the setpoints are
displayed.
ALARM KEY
DIGIT SELECT KEY
VALUE SELECT KEY
Alarm value blinks and Alarm
1 LED indicator lights. Press
to select digit.
Using
to select digit and
to set digit value, enter
setpoint 1 value.
Press the key to
display Alarm 1 value.
Press key again to
display Alarm 2 value
Alarm value blinks and Alarm
2 LED indicator lights. Press
to select digit.
Using
to select digit and
to set digit value, enter
setpoint 2 value.
Press key again. The meter resets and then displays the present reading.
ALARMS
ALARMS
ALARMS
-35-