Laurel Electronics LAUREATE SERIES DPM User Manual
Page 31
-29-
Press the
key to display
dEcPt (Decimal point).
Press
to display decimal
point location.
Use to set digit values and
set to .00000 ( 0mV ).
Note: Most significant digit
may be set to 0 thru 9 and
-0 thru -9.
Press the
key to display
Lo in (Low signal input value).
Press
to display value.
Press
again to select
another digit. Note: Deci-
mal point is fixed by input
range selection.
Press
to display value.
Note: Decimal point is fixed
by dEcPt selection.
Use to set digit values and
set to 000.00
Use to set digit values and
set to .02000 ( 20mV )
Press the
key to display
Hi in (High signal input value).
Press
to display value.
Note: Decimal point is fixed
by input range selection.
Use to set digit values and
set to 100.00
Press
to display value.
Note: Decimal point is fixed
by dEcPt selection.
Press the
key to display
Hi rd (Desired meter reading
at high signal input).
DIGIT SELECT KEY
VALUE SELECT KEY
Press
to select decimal
point location.
Press the
key. Continue to press
(or
and simultaneously) until rESEt is
displayed. The meter will now go to the operating mode and display the value of the input signal.
1 2 3 4 5
Press the
key to display
SEtuP. (Basic setup). See
Section 9, Page 10 for de-
tailed description of selec-
tions for digits 1 through 5.
Press
to display status.
Press
again to select
another digit. Selected digit
will flash.
Press to select value.
Digit 1:"0"= 20,000 cts. FS
"2"=LSD fixed zero
"3"=2,000 cts. FS
Digit 4: "1" for 2 point scaling
Press the
key to display
Lo rd (Desired meter reading
at low signal input).
MENU KEY