14.3
MENU SELECTIONS
The following example is setup for a 4-wire DIN RTD. When setting up the meter, it may be
necessary to enable some of the menu items. See Section 9 for further information.
DIGIT SELECT KEY
VALUE SELECT KEY
Press the
key to display
InPut (Input type selection).
Note: Selection of input type
& range must match jumper
selection in Section 14.1.
Press
until rtd
(resistance temperature
detector) is displayed
Press
to select rtd type
(4d°F, 4d°C, 4A°F, 4A°C,
3d°F, 3d°C, 3A°F, 3A°C,
2d°F, 2d°C, 2A°F, 2A°C)
Number = # of leads
Letter = Din or ANSI RTD
Press
to display status
and select left digit. Press
again to select another
digit. Selected digit will flash.
Press to select value for
flashing digit.
Digit 1:
"0"= 0.1 degree resolution
"2" = 0.01 degree resolution
"3"=1 degree resolution
1 2 3 4 5
Press the
key to display
SCALE. Divide 100 by
resistance of RTD at 0° C to
calculate scale factor.
Press
to display value
and select left digit. Press
again to select another digit.
Use
to set digit values.
Set value and decimal to cal-
culated scale factor.
Press the
key to display
OFFSt (Zero offset).
Press
to display value
and select left digit. Press
again to select another
digit.
Use to set digit values and
set to 0.0 for °F and °C or
set value to 273.2 if °C is
selected to display in °Kelvin
and 459.7 if °F is selected to
display in ° Rankine.
Press the
key to display
SEtuP. (Basic setup)
See Section 9, Page 10 for
detailed description of selec-
tions for digits 1 through 5.
MENU KEY
-27-
Press the
key. Continue to press
(or
and simultaneously) until rESEt
is displayed. The meter will go to the operating mode and display the value of the input signal.