beautypg.com

Laurel Electronics LAUREATE SERIES DPM User Manual

Page 30

background image

-28-

Press

until rAtio (Ratio-

metric operation) is dis-
played.

15

STRAIN GAUGES AND POTENTIOMETERS

DIGIT SELECT KEY

VALUE SELECT KEY

Press the

key to display

InPut. Note: Range selec-
tion must match jumper se-
lection in Section 15. 1 .

Press to select 0.2U, 2.0U
or 20.0U (200mV, 2V or
20Vdc).

MENU KEY

15. 2 MENU SELECTION

12.1

RANGE JUMPER SELECTIONS

Voltage
Input Jumpers Required

200mV

E, b

2V

E, a

20V

F, g, b

200V

F, g, a

660V

F, h, a

Current
Input Jumpers Required

2mA

D, h, b

20mA

C, h, b

200mA

B, h, b

5A

A, h, b

a

b

A

B

C

D

E

F

g

h

Voltage

Input

E1

E2 E3

±200.00 mV

A

f

b

±2.0000 V

A

f

a

±20.000 V

B

h

b

±200.00 V

B

h

a

±600.00 V

B

g

a

Current

FS Input

E1

E2 E3

±2.0000 mA

A

e, h

b

±20.000 mA

A

d, h

b

±200.00 mA

A

c, h

b

±5.000 A

A a,b,h

b