4 system and warning alarm table, Table 6, Lists – Hach-Lange POLYMETRON 9245 User Manual User Manual
Page 81: User setup 7.3.4 system and warning alarm table

79
User Setup
7.3.4 System and warning alarm table
The following table lists the different system and warning alarms:
Table 6 System and warning alarms
Message
Description
Category
Reset measure mod.
Reset the measure module
System
Reset LC module
Reset the local controller module
System
LC Error
Local controller module error
Warning
Err. measure mod.
Measure module error
Warning
Comms error
Communication error
Warning
Temp > 45°C (113°F)
Temperature too high
Warning
Temp < 5°C (45°F)
Temperature too low
Warning
Measure error
Measurement error
Warning
Meas.time over
Measurement cycle time too long
Warning
Cal.time over
Calibration cycle time too long
Warning
Flow x < min
Not enough sample
Warning
No condit.
No sample conditioner (< 50 mL)
System
Channel inhib.
The sample measurement is by-passed
Warning
Cal. sol < 100 mL
Calibration solution level low
Warning
Electro. < 10 mL
Electrolyte volume level low
Warning
Activ. sol. < 100 mL
Reactivation solution level low
Warning
Condit. < 100 mL
Conditioning solution level very low
Warning
P0 too high
Base measurement P0 is too high
Warning
P2 < 5 x P1
2nd point calibration < 5 times 1st point
Warning
Slope drift
Slope over tolerance
Warning
Offset drift
Offset over tolerance
Warning
Stabilize error
Stability criteria for accuracy not reached
Warning
Sample x < C
Sample concentration x is below threshold C
Warning
Sample x < 0 ppb
Sample concentration x is negative
Warning
Sample x > 1000 ppm
Sample concentration is too high
Warning