Analyzer Overview
3.10.4 Auto calibration process
3.10.5 Manual calibration process
3.11 Available options
3.11.1 K-Kit (cationic)
For a high acidity water such as that from a cation exchanger outlet, the regular gaseous
conditioning is not sufficient to raise the pH to values superior to 10.3. The forced-gas
conditioning system (K-kit) is then needed.
The K-Kit option includes a gas pump with no moving parts, a power supply board, additional
hydraulics and full installation instructions.
For further information contact your local Hach Lange representative.
3.11.2 Static heat exchanger system
A static heat exchanger system is available as an option. It comes complete with inlet and outlet
connectors (4/6mm tubing) and mounting (2 flanges and screws).
Very easy to install, and requiring no voltage supply, this compact (350 x 40mm) product
absorbs changes of heat even on samples flowing at 5 L/h per channel.
Specially designed for POWER applications, it has a high resistance to corrosion and deposits,
and allows incoming samples from 0 - 60°C to be released to the analyzer within its operating
range of 5 - 45°C.
For further information contact your local Hach Lange representative.
Action
Screen
4-20 mA
Alarms
RS485
Conc.
Other prog.
Address 0
Others
Phase 1
Actual
Actual
Event
Calibration
N/A
Actual
Address 163=10
Phase 2
Frozen
Frozen
Frozen
Address 163=13
Phase 3
Phase 4
Address 163=10
Phase 5
Actual
Actual
Actual
Phase 6
Frozen
Frozen
Frozen
Phase 7
Phase 8
Actual
Actual
Actual
Phase 9
Frozen
Frozen
Frozen
Action
Screen
4-20 mA
Alarms
RS485
Conc.
Other prog.
Address 0
Others
Phase 1
Frozen
Frozen
Event
Calibration
N/A
Frozen
Address 163=13
Phase 2
Address 163=11
Phase 3
Phase 4
Phase 5
Actual
Actual
Actual
Phase 6
Frozen
Frozen
Frozen
Phase 7
Phase 8
Actual
Actual
Actual