Operating instructions – Hach-Lange POLYMETRON 9245 User Manual User Manual
Page 63

61
Operating Instructions
Upon completion of the rinsing step, the analyzer is ready for 
the first manual step in the cycle. 
Place the lid of the overflow vessel on the side and pour 
approximately 200 mL of the solution into the overflow vessel, 
as prompted on screen. Manual introduction is complete when 
the sample overflows at the back of the overflow vessel. 
Place the lid back on top of the overflow vessel and select OK 
to start the measurement.
The display switches back to the main measurement screen 
and displays the measurement value.
The progress bar at the top of the screen monitors the 
measurement time of the sample.
If you wish to abort the process at any time during the 
measurement cycle, select Stop.
Once measurement of the solution is complete, the measured 
value and time will be displayed against verif. s. This should 
be almost identical to the value entered in the VALID. SOL 
field previously.
The value displayed in gap, will be the percentage difference 
between the concentration of the verification solution entered 
in the VALID. SOL field, and the concentration measured by 
the instrument. If this value is too high, then a calibration of 
the analyzer is recommended.
Once the cycle is complete, the analyzer automatically reverts 
back to standard measurement mode.
Select Main to go back to the main measurement screen.