25

ENGLISH

4.11. Data Logger

The si629 C is able to store up to 200 data in memory. When the data logger is full, data will be stored
replacing the eldest ones. Data can be stored following two different criteria:

Interval: data are stored with the frequency programmed in Interval:
XX min.

Event: data are stored only if the values are higher o lower than the
programmed value, and with the selected frequency.

It displays all stored data. The first record will be
displayed, and then it will automatically go on
one by one every 1 sec.
Press

to quit the View display.

Press

to stop the automatic view. The display

will be frozen at the last displayed record.
Press to see data in increasing order.
Press to see data in decreasing order.

DATA LOGGER

CODE 000

MEASURING

22.3°C

545 µS/cm @25°

D. LOG.

> View

Program
Erase

MEASURE

MEASURING 22.3°C

545 µS/cm @25°

Press

to select the desired option.

Select START to start the storage
of data.
Select STOP to end this option.

To empty the
data logger
select the option
erase.

VIEW

> Max.

Min.
All

Min 12-04 / 11:21
520 µS/cm 25.2°C

Event
EC

t= 30 m

Max 12-04 / 12:22
1250 µS/cm 25.2°C

Intervalo

t= 30 m

001 12-04 / 11:01

545 µS/cm 25.2°C

Date
10-03-2004

PROG.

> Interval *

Event

Start

Stop *

Date

Hoor

Hour
12:04:56

ERASE

DATA LOGGER
ERASE

Press

to select the

desired option.

Select the value and the unit (sec., min. or hours)

It displays only the maximum stored reading.

It displays only the minimum stored reading.

4. Operation

Press

until the desired option is reached.

Enter code 100 and y accept with

si629 C_HACH LANGE_Ed 0510.indd ANG:25

si629 C_HACH LANGE_Ed 0510.indd ANG:25

20/5/10 12:06:58