Terminal assignment (interior panel), Danger, Installation english – Hach-Lange SI629 C User Manual

Page 13

background image

9

3. Installation

ENGLISH

Terminal assignment (interior panel)

Important notes:

• El si629 C does not require ground conec-

tion (3 or 4 terminals).

• In environnements with interference prob-

lems it is recommended to connect the in-
strument to ground
in terminal 3.

If ground connection is not available, it is

recommended to leave terminal 3 free and
connect the additional shield of the triaxial
cable to the terminal 8 (Pt 1000 shielding).

L 1

A

L 2 / C

4-20 mA

CELL

VCA

Pt 1000

WHEN IT IS ON, IT INDICATES A FUSED FUSE.

DANGER

Relays:
- Limit the current up to 3 A by using one fuse.
- Do not connect cables with voltage higher than 24 V.

si629 C_HACH LANGE_Ed 0510.indd ANG:9

si629 C_HACH LANGE_Ed 0510.indd ANG:9

20/5/10 12:06:50

20/5/10 12:06:50