1-2

Brooks

®

4800 Series

Section 1 Introduction

Installation and Operation Manual

X-TMF-4800-MFC-eng

Part Number: 541B072AAG

September, 2013

Do not operate on gases other than those listed in this
specification. Failure to heed this warning can result in serious
personal injury and/or damage to the equipment.

WARNING

!

Operating Limits
Pressure: 0 – 10 barg
Temperature: 0 - 50°C at 10 barg
Operating Humidity: 5 to 95% R.H. (ambient)

Pressure Differential Range (Controllers)
Minimum: 0.35 bar (5 psid)
Maximum: 10 bar (150 psid)

Leak Integrity

Inboard to Outboard: 1x10

-9

atm scc/sec Helium max.

ELECTRICAL CHARACTERISTICS:

Electrical Connections
Analog/RS232:15-pin D-sub connector
(pin-out schematic shown in Table 1-1)
All pins shall be protected against ESD (Electro Static Discharge)
All input pins are protected from electrical stress, to at least withstand 28
Vdc (power supply)

Power Supply Voltage**
+15 Vdc + 10% or
+24 Vdc + 10%
Device only uses single sided power supply
Inrush current: <1 A
Power suplly is protected from reverse polarity

Power Requirements

15 Vdc

24 Vdc

Device

Typical

(mA)

Max.
(mA)

Typical

(mA)

Max

(mA)

4860 (without valve)

30

60

30

60

4850 (with valve)

130

160

150

200

Analog Input/Output
Command/Setpoint Input

Voltage: 0-5 Vdc, minimum impedance 990

Current: 4-20 mA, maximum impedance 220

Flow Signal Output

Voltage: 0-5 Vdc, minimum load 2000

Current: 4-20 mA, maximum load 550

* All performance characteristics are at calibration conditions.
** For high flows or low differential pressures (using orifices 0.049" (1.25mm) or

0.079" (2.0mm)) only 24 Vdc power is available.