13.2.3. Command mode

In the command mode, the scale is controlled by commands that come from an
external device such as a personal computer.

Command List

Command Description

Remarks

Q

Send data immediately.

Z

Zero the scale when the weight is stable.

Same as the ZERO switch.

T

Tare the scale when the data is stable.

Same as the TARE switch.

U

Switch the weighing unit.

Same as the UNITS switch.

?H2

When the five-level comparator mode is used:
Send the current HIHI limit value.
When the three-level comparator mode is used:
Send the current HI limit value.

?H1

When the five-level comparator mode is used:
Send the current HI limit value.
When the three-level comparator mode is used:
Not used

?L1

When the five-level comparator mode is used:
Send the current LO limit value.
When the three-level comparator mode is used:
Not used

?L2

When the five-level comparator mode is used:
Send the current LOLO limit value.
When the three-level comparator mode is used:
Send the current LO limit value.

Send a setting value.

Function settings

Five-level (

Cp-l 0

)

Three-level (

Cp-l 1

)

H2

When the five-level comparator mode is used:
Set the HIHI limit value.
When the three-level comparator mode is used:
Set the HI limit value.

H1

When the five-level comparator mode is used:
Set the HI limit value.
When the three-level comparator mode is used:
Not used.

L1

When the five-level comparator mode is used:
Set the LO limit value.
When the three-level comparator mode is used:
Not used.

L2

When the five-level comparator mode is used:
Set the LOLO limit value.
When the three-level comparator mode is used:
Set the LO limit value.

Set the six-digit value excluding