Function list – A&D Super Wash Down Scales SW-150KL User Manual
Page 25
23
12.3. Function list
Item
Parameter
Description
0
Auto power-off disabled
Auto power-off function
poff
1
Auto power-off enabled
Turns the power OFF
automatically
0
Normal (1/3,000)
1
High (1/6,000 or 1/7,500)
Weight display
resolution
re5o
2
Maximum (1/12,000 or 1/15,0 00)
Changes the
minimum display
0
kg
Weighing unit
Ut-g
1
g
0
Zero tracking function disabled
Zero tracking
trc
1
Zero tracking function enabled
Tracks the zero drift
0
±0.5d (width 1d)
1
±1.0d (width 2d)
Stability band
width
5t-b
2
±2.0d (width 4d)
0
0.5 second
1
1.0 second
Stability
detection time
5t-t
2
2.0 seconds
Conditions to turn the
STABLE indicator on
d = minimum display
in kg
0
Weak stability / fast response
1
Normal stability / normal response
Stability
Response speed
Cond
2
Strong stability / slow response
Filtering
0
2400 bps
1
4800 bps
Serial interface
baud rate
bp5
2
9600 bps
0
7 bits / EVEN
1
7 bits / ODD
Serial interface
data bits / parity
btpr
2
8 bits / Non parity
0
Stream mode / Command mode
1
Command mode only
2
Print switch mode /
Command mode
3
Auto-print mode +/- data / Command mode
Serial interface
Data output
mode
prt
4
Auto-print mode + data/ Command mode
0
RS-232C
1
RS-422
Serial interface
Type
5if
2
RS-485
Serial interface
Address
adr
##
## = 00 to 99
(Factory setting: ## = 01)
0
No reply except data to commands
Serial interface
Response
aCk
1
Reply to commands
Only when the
options, RS-232C or
RS-422/485, are
used.
0
Five-level (LOLO, LO, OK, HI, HIHI)
Comparator mode
Cp-l
1
Three-level (LO, OK, HI)