Restoring the factory set values – A&D Super Wash Down Scales SW-150KL User Manual
Page 23
21
11.3. Restoring the factory set values
If the gravity acceleration value or calibration weight value is changed unintentionally,
restore those values to the factory setting values, as follows.
Make sure that the scale is in the weighing
mode (displaying “kg” (“g”), “lb”, “oz” or “pcs”).
1. Press and hold the TARE switch until
the gravity acceleration value appears, and
release the switch.
2. Press the UNITS switch twice to
display “
ClrC
” .
.
3. Press the TARE switch to display
“
ClrC no
” with “
no
” blinking.
4. Press the PRINT switch.
“
ClrC no
” changes to “
ClrC go
” with
“
go
” blinking.
5. When “
ClrC go
” is displayed, press the
TARE switch. The factory setting values
are restored and “
end
” is displayed.
To finish the restoring procedure, turn the
power OFF.
To cancel the restoring procedure, press the
ZERO switch. The display returns to step 2.
Press and hold
(more than 4 seconds)
Release
Press twice
Press
Press
Press