Calibration, Gravity acceleration correction – A&D Super Wash Down Scales SW-150KL User Manual
Page 21
19
11. CALIBRATION
Adjusts the scale for accurate weighing.
Calibrate the scale in the following cases.
When the scale is first installed.
When the scale has been moved.
When the ambient environment has changed.
For regular calibration.
11.1. Gravity acceleration correction
When the scale is first used or has been moved to another location, it should be
calibrated using a calibration weight.
But if a calibration weight is not available, the gravity acceleration correction will
compensate the scale. Change the gravity acceleration value stored in the scale to the
value of the area where the scale will be used. Refer to the gravity acceleration map at
the end of this manual.
Note: Gravity acceleration correction is not required when the scale will be
calibrated using a calibration weight at the place where it is to be used.
Make sure that the scale is in the weighing
mode (displaying “kg” (“g”), “lb”, “oz” or “pcs”).
1. Press and hold the TARE switch until the
gravity acceleration value (in this example:
9.7985) appears, and release the switch.
2. Press the TARE switch to enter the
gravity acceleration value setting mode.
3. Change the displayed value using the following
switches.
UNITS To shift the blinking digit to the
right.
PRINT To increase the value of the
blinking digit by one.
4. Press the TARE switch. The setting value
is stored and “
end
” is displayed.
5. When calibration using a calibration weight is
to be performed, go to step 3 of “11.2.
Calibration using a weight”. To finish the
setting procedure, turn the power OFF.
Press and hold
(more than 4 seconds)
Release
Press
Press
Press