background image

14 IOM-DPT-A: FEB 2014

5.4. Operating conditions.

Ambient temperature range

-40°C

85°C

Operating temperature for intrinsic-safe versions

(Exi) in

accordance with „Appendix Exi”.

Operating temperature for Exd versions

in accordance

with „Appendix Exd.01”.

Relative humidity

up to 98%

Medium temperature range

Pt100

- 200 ... 550 °C

N-Cr-NiAl - 40 ... 550 °

C

Thermal compensation range

-

25 … 75

0

C

5.4.1. Electromagnetic Compatibility (EMC), immunity

Rating according to EN 61326-1,2for industrial applications

Electrostatic Discharge Immunity (ESD):
EN 61000-4-2; S3 level: contact ±6kV, air ±8kV; criterion A

Conducted Radio Frequency:
EN 61000-4-

6; 0,15… 80MHz, 10V; criterion A


Radiated electromagnetic Field:
EN 61000-4-

3; 80… 2 000MHz – 10V/m, 2 000 … 2 700MHz – 1V/m; criterion A


Electrical Fast Transient (Burst Immunity)
EN 61000-4-4; ± 2kV power supply port/earth, ± 1kV signal port/earth; criterion A

Electrical Slow Transient (Surge Immunity):
EN 61000-4-5; ±0.5kV (±1kV) 0,5kV differentia mode, 1kV common mode; criterion B

5.4.2. Electromagnetic Compatibility, emission:

According to CISPR16-1, CISPR 16-2, class B, distance to the antenna 3m, quasi-peak measuring:
Radiated emission:

0,15 … 30MHz, 80-52dBμV/m;

30 … 2000MHz, <54dBμV/m


Conducted emission:

0,01 … 0,150MHz, 96-50dBμV/m;

0,150 … 0,350MHz, 60-50dBμV/m;

0,35 … 30MHz, <50dBμV/m

5.4.3. Climatic immunity: dry heat, cold, humidity, salt mist:

Dry heat:
EN 60068-2-2, test B; T = 70

0

C, RH = max 55%


Cold :
EN 60068-2-1, test A; T = -25

0

C,


Damp heat cycle:
EN 60068-2-30, test D ; (T = 55

0

C, RH = min95%, 24h)x2

5.4.4. Mechanical immunity

Shocks:
EN 60068-2-27, 50g/11ms

Vibrations:
EN 60068-2-6, test Fc; up to

1,6mm for 2 … 25Hz, up to 4g for 25 … 100Hz