1) The MAXimum amplitude is dependent on the offset.
2) The MAX and MIN arguments should not be used in a
program message containing an OFFSet command, since
these values are evaluated during parsing, based on the
current value of the offset.


OFFSet

The offset command is used to set the DC offset of the
output waveform. Note that the amplitude and the offset
are limited by the relation:

Peak Amplitude + |Offset| ≤ 5V


Arguments
Type:

Numeric

Units:

V, mV

Range:

10mV to 5V

Rounding: To 10mV
Command Type: Setting or Query
Setting
Syntax:

OFFSet[units]

OFFSetMINimum|MAXimum
Examples:

OFFS 2.5

OFFS 2.5V
OFFS MAX
Query
Syntax:

OFFSet?[MINimum|MAXimum]

Examples:

OFFS?

OFFS? MAX