Data Sheet
Pinpoint
®
Trigger System
Sensitivity
DPO and MSO Models
DSA Models
Internal DC Coupled
4% of full scale from DC to 50 MHz
10% of full scale at 4 GHz
20% of full scale at 8 GHz
50% of full scale at 11 GHz
External (Auxiliary Input) 50 Ω
250 mV from DC to 50 MHz, increasing to 350 mV at 1.0 GHz
Trigger Characteristics
A Event and Delayed B Event Trigger Types
Edge, Glitch, Runt, Width, Transition Time, Time-out, Pattern, State, Setup/Hold, Window – all except
Edge, Pattern, and State can be Logic State qualified by up to two channels
Main Trigger Modes
Auto, Normal, and Single
Trigger Sequences
Main, Delayed by Time, Delayed by Events, Reset by Time, Reset by State, Reset by Transition. All sequences
can include separate horizontal delay after the trigger event to position the acquisition window in time
Clock Recovery System
Requires Option PTH or Option MTH
Standard
Clock Recovery Phase Locked Loop Bandwidth
Fixed at FBaud/1600
Clock Recovery Frequency Range
1.5 MBaud to 3.125 GBaud
Requires Option PTU
Requires Option STU
8b10b Max Baud Rate
5 GBaud
Requires Option MTH
Standard
Communications-related Triggers
Support for AMI, HDB3, BnZS, CMI, MLT3 and NRZ encoded communications signals. Select among isolated
positive or negative one, zero pulse form or eye patterns as applicable to the standard
Requires Option PTH
Standard
Up to 64 bit serial word recognizer, bits specified in binary (high, low, don’t care) or hex format.
Trigger on NRZ-encoded data up to 1.25 GBaud
Trigger on 8b/10b-encoded data from 1.25 to 3.125 GBaud (40 bits)
Requires Option PTU
Requires Option STU
Serial Pattern Trigger
Trigger on 8b/10b-encoded data up to 5 GBaud (40 bits)
Threshold Range
P6780
–2 to +4.5 V
P6717
–1.5 to +4 V
Threshold Accuracy
±100 mV + 3% of threshold setting
Maximum I
2
C, SPI Bus Toggle Rate
10 MHz
Clock Recovery Jitter (RMS)
<0.25% bit period + 2 ps
RMS
for PRBS data patterns
<0.25% bit period + 1.5 ps
RMS
for repeating "0011” data pattern
Enhanced Triggering
User-selectable; enhanced triggering corrects the difference in timing between the trigger
path and the acquired data path (supports all Pinpoint trigger types on both A- and B-Events
except pattern trigger); Not available in FastAcq mode
Line
Trigger on power line signal. Level fixed at 0 V
Minimum Signal Amplitude needed for Clock Recovery
1 div
p-p
up to 1.25 Gbaud
1.5 div
p-p
above 1.25 Gbaud
Tracking/Acquisition Range
±2% of requested baud
Trigger Coupling
DC, AC (attenuates <100 Hz)
HF Rej (attenuates >20 kHz)
LF Rej (attenuates <200 kHz)
Noise Reject (reduces sensitivity)
Trigger Holdoff Range
250 ns min to 12 s max
Trigger Level Range Internal
±120% of full scale from center of screen
18