Pinpoint

®

Trigger System

Sensitivity

DPO and MSO Models

DSA Models

Internal DC Coupled

4% of full scale from DC to 50 MHz

10% of full scale at 4 GHz

20% of full scale at 8 GHz

50% of full scale at 11 GHz

External (Auxiliary Input) 50 Ω

250 mV from DC to 50 MHz, increasing to 350 mV at 1.0 GHz

Trigger Characteristics

A Event and Delayed B Event Trigger Types

Edge, Glitch, Runt, Width, Transition Time, Time-out, Pattern, State, Setup/Hold, Window – all except

Edge, Pattern, and State can be Logic State qualified by up to two channels

Main Trigger Modes

Auto, Normal, and Single

Trigger Sequences

Main, Delayed by Time, Delayed by Events, Reset by Time, Reset by State, Reset by Transition. All sequences

can include separate horizontal delay after the trigger event to position the acquisition window in time

Clock Recovery System

Requires Option PTH or Option MTH

Standard

Clock Recovery Phase Locked Loop Bandwidth

Fixed at FBaud/1600

Clock Recovery Frequency Range

1.5 MBaud to 3.125 GBaud

Requires Option PTU

Requires Option STU

8b10b Max Baud Rate

5 GBaud

Requires Option MTH

Standard

Communications-related Triggers

Support for AMI, HDB3, BnZS, CMI, MLT3 and NRZ encoded communications signals. Select among isolated

positive or negative one, zero pulse form or eye patterns as applicable to the standard

Requires Option PTH

Standard

Up to 64 bit serial word recognizer, bits specified in binary (high, low, don’t care) or hex format.

Trigger on NRZ-encoded data up to 1.25 GBaud

Trigger on 8b/10b-encoded data from 1.25 to 3.125 GBaud (40 bits)

Requires Option PTU

Requires Option STU

Serial Pattern Trigger

Trigger on 8b/10b-encoded data up to 5 GBaud (40 bits)

Threshold Range

P6780

–2 to +4.5 V

P6717

–1.5 to +4 V

Threshold Accuracy

±100 mV + 3% of threshold setting

Maximum I

2

C, SPI Bus Toggle Rate

10 MHz

Clock Recovery Jitter (RMS)

<0.25% bit period + 2 ps

RMS

for PRBS data patterns

<0.25% bit period + 1.5 ps

RMS

for repeating "0011” data pattern

Enhanced Triggering

User-selectable; enhanced triggering corrects the difference in timing between the trigger

path and the acquired data path (supports all Pinpoint trigger types on both A- and B-Events

except pattern trigger); Not available in FastAcq mode

Line

Trigger on power line signal. Level fixed at 0 V

Minimum Signal Amplitude needed for Clock Recovery

1 div

p-p

up to 1.25 Gbaud

1.5 div

p-p

above 1.25 Gbaud

Tracking/Acquisition Range

±2% of requested baud

Trigger Coupling

DC, AC (attenuates <100 Hz)

HF Rej (attenuates >20 kHz)

LF Rej (attenuates <200 kHz)

Noise Reject (reduces sensitivity)

Trigger Holdoff Range

250 ns min to 12 s max

Trigger Level Range Internal

±120% of full scale from center of screen

18