Digital phosphor oscilloscopes – Atec Tektronix-TDS7000B Series User Manual
Page 6
Digital Phosphor Oscilloscopes
TDS7000B Series
Trigger Modes
Edge – Positive and/or negative slope on any
channel or front panel auxiliary input. Coupling
includes DC, AC, noise reject, HF reject and
LF reject.
Glitch – Trigger on or reject glitches of positive,
negative or either polarity. Minimum glitch width is
1.0 ns with 200 ps resolution (TDS7104/TDS7054).
Minimum glitch width is 170 ps (TDS7704B) or
225 ps (all other B models) with rearm time of
250 ps (B models).
Width – Trigger on width of positive or negative
pulse (down to 170 ps on B models) either within
or out of selectable time limits – 1 ns.
(TDS7104/TDS7054) or 340 ps (B models) to 1 s).
Runt – Trigger on a pulse that crosses one
threshold but fails to cross a second threshold before
crossing the first again. Optional time qualification.
Timeout – Trigger on an event which remains high,
low or either, for a specified time period, selectable
from 1 ns (TDS7104/TDS7054) or 340 ps (B models)
to 1 s with 200 ps (TDS7104/TDS7054) or 100 ps
(B models).
Transition – Trigger on pulse edge rates that are
faster or slower than specified. Slope may be positive,
negative or either.
Setup/Hold – Trigger on violations of both setup
time and hold time between clock and data present
on any two input channels.
Pattern – Trigger when pattern goes false or stays
true for specified period of time. Pattern (AND, OR,
NAND, NOR) specified for four input channels
defined as HIGH, LOW or Don’t Care.
State – Any logical pattern of channels (1, 2, 3)
clocked by edge on channel 4. Trigger on rising
or falling clock edge.
Window – Trigger on an event that enters or exits a
window defined by two user-adjustable thresholds.
Event can be time or logic qualified (B models only)
Logic Qualified Trigger applicable to Glitch, Width,
Runt, Timeout, Transition, Setup/Hold, Window trig-
gers – trigger on the specified event only if the logic
state defined with the remaining unused channels
occurs (B models only).
Trigger Delay by Time –
16 ns (5 ns for B models) to 250 s.
Trigger Delay by Events –
1 to 10,000,000 Events.
Digital Phosphor Oscilloscopes • www.tektronix.com/oscilloscopes
6
Clock Recovery System (Option SM, ST)
TDS7154B/TDS7254B/TDS7404B/TDS7704B
Clock Recovery Phase Locked Loop Bandwidth
Fbaud/1600 typical
Tracking/Acquisition Range
±2% of requested baud
Clock Recovery Jitter (Typical)
0.25% period +5 ps
RMS
for PRBS data pattern
or 4 ps
RMS
for repeating “011” data patterns
Input Sensitivity for Clock Recovery
1 division peak-to-peak displayed signal
Input Data Rates
1.5 Mbaud to 3.125 Gbaud