beautypg.com

Digital phosphor oscilloscopes – Atec Tektronix-TDS7000B Series User Manual

Page 4

background image

Digital Phosphor Oscilloscopes

TDS7000B Series

Digital Phosphor Oscilloscopes • www.tektronix.com/oscilloscopes

4

Time Base System

TDS7054/TDS7104

TDS7154B/TDS7254B/

TDS7404B/TDS7704B

Time Base Range

200 ps/div to 40 s/div

50 ps to 10 s/div

Time Base Delay Time Range

16 ns to 250 s

5 ns to 250 s

Channel-to-Channel Deskew Range

±25 ns

±75 ns

Delta Time Measurement Accuracy

±((0.06/sample rate) + (15 ppm x reading)) RMS

±((0.06/sample rate) + (2.5 ppm x reading)) RMS

Trigger Jitter (RMS)

8 ps

RMS

(typical)

2 ps

RMS

(typical) (7254B/7154B)

1.5 ps

RMS

(typical) (7404B)

1.2 ps

RMS

(typical) (7704B)

Long Term Sample Rate and Delay

±15 ppm over

≥1 ms interval

2.5 ppm over any

≥100 ms interval

Time Accuracy

Acquisition System

TDS7054

TDS7104

TDS7154B/ TDS7254B/

TDS7404B/ TDS7704B

Real-time Sample Rates

1 channel (max)

5 GS/s

10 GS/s

20 GS/s

2 channels (max)

5 GS/s

5 GS/s

10 GS/s

3-4 channels (max)

2.5 GS/s

2.5 GS/s

5 GS/s

Equivalent Time Sample Rate (max)

250 GS/s

250 GS/s

1 TS/s

Maximum Record Length per Channel

2 Mb (1-CH.), 1 Mb (2-CH.), 500 Kb (4-CH.)

4 Mb (1-CH.), 2 Mb (2-CH.),

with Standard Memory

1 Mb (4-CH.)

With Memory Opt. 2M

8 Mb (1-CH.), 4 Mb (2-CH.),

2 Mb (4-CH.)

With Memory Opt. 3M

16 Mb (1-CH.), 8 Mb (2-CH.),

4 Mb (4-CH.)

With Memory Opt. 4M

32 Mb (1-CH.), 16 Mb (2-CH.),

8 Mb (4-CH.)

With Memory Opt. 5M

64 Mb (1-CH.), 32 Mb (2-CH.),

16 Mb (4-CH.)

Maximum Duration at Highest Real-time Resolution (1-CH)

TDS7054

TDS7104

TDS7154B/TDS7254B/

TDS7404B/TDS7704B

Time Resolution (Single-shot)

200 ps (5 GS/s)

100 ps (10 GS/s)

50 ps (20 GS/s)

Max Duration with Standard Memory

400 µs

200 µs

200 µs

Max Duration with Opt. 2M

1.6 ms

800 µs

400 µs

Max Duration with Opt. 3M

3.2 ms

1.6 ms

800 µs

Max Duration with Opt. 4M

1.6 ms

Max Duration with Opt. 5M

3.2 ms