Characteristics, Digital phosphor oscilloscopes – Atec Tektronix-TDS7000B Series User Manual
Page 3
Digital Phosphor Oscilloscopes
TDS7000B Series
Digital Phosphor Oscilloscopes • www.tektronix.com/oscilloscopes
3
Vertical System
TDS7054
TDS7104
TDS7154B
TDS7254B
TDS7404B
TDS7704B
Input Channels
4
4
4
4
4
4
Hardware Analog
500 MHz
1 GHz
1.5 GHz
*1
2.5 GHz
*1
4 GHz
*1
7.25 GHz
*1
Bandwidth (–3 dB)
(typical), 7GHz
*1
guaranteed
Rise Time 10%
800 ps
400 ps
200 ps
130 ps
100 ps
62 ps
to 90% (typical)
Rise Time 20%
135 ps
83 ps
72 ps
43 ps
to 80% (typical)
DC Gain Accuracy
1%
±(2% +
±2.5% +
(2% x offset)
(2% x offset)
Hardware Bandwidth
250 MHz or 20 MHz
Requires TCA-1 MEG
Limits
Input Coupling
AC, DC, GND
DC, GND
Input Impedance
1 M
Ω ±0.5% or 50 Ω ±1%
50
Ω ±2.5%
Input Sensitivity, 1 M
Ω
1 mV/div to 10 V/div
—
Input Sensitivity, 50
Ω
1 mV/div to 1 V/div
2 mV/div to 1 V/div
Vertical Resolution
8-Bit (>11-Bit with averaging)
8-Bit (>11-Bit with averaging)
Max Input Voltage,
±150 V CAT I Derate at 20 dB/decade
—
1 M
Ω
to 9 V
RMS
above 200 kHz
Max Input Voltage,
5 V
RMS
, with peaks less than ±30 Volts
<1 V
RMS
for <100 mV/div, <5 V
RMS
for
≥100 mV/div settings
50
Ω
Also determined by TekConnect
®
accessory
Offset Range
1 mV/div to 100 mV/div: ±1 V
2 mV to 50 mV/div: ±0.5 V
101 mV/div to 1 V/div: ±10 V
50.5 mV to 99.5 mV: ±0.25 V
1.01 V/div to 10 V/div: ±100 V
100 mV to 500 mV: ±5 V
505 mV to 1 V/div: ±2.5 V
Channel-to-Channel
≥100:1 at 100 MHz and
≥80:1 at 1.5 GHz and
Isolation Any Two
≥30:1 at the rated bandwidth
≥15:1 at the rated bandwidth
Channels at Equal
Vertical Scale Settings
≥ ≥≥ ≥
*1
At
≥10 mV/div.
Note: Typical system bandwidth of TDS7404B with P7240: 4 GHz.
Note: Typical system bandwidth of TDS7404B with P7330: 3.5 GHz.
Note: Typical system bandwidth of TDS7704B with P7260: 6 GHz.
Note: Typical system bandwidth of TDS7704B with P7350: 5 GHz.
Note: Typical system bandwidth of TDS7704B with P7380: 7 GHz.
Characteristics