Ohms Measurement

Accuracy

±

Ohms Range

4-Wire

2- and 3-Wire*

0 to 400

0.1

0.15

400 to 1.5 k

0.5

1.0

1.5 to 3.2 k

1

1.5

Excitation Current: 0.2 mA
Maximum input voltage: 30 V
Temperature coefficient -10

°

C to 18

°

C, +28

°

C to

55

°

C:

±

0.005 % of range per

°

C

* 2-wire: Does not include lead resistance.

3-wire: Assumes matched leads with a total resistance
not exceeding 100

.

Ohms Source

Ohms Range

Excitation Current from

Measurement Device

Accuracy

±

15 to 400

0.15 to 0.5 mA

0.15

15 to 400

0.5 to 2 mA

0.1

400 to 1.5 k

0.05 to 0.8 mA

0.5

1.5 to 3.2 k

0.05 to 0.4 mA

1

Resolution

15 to 400

0.1

400 to 3.2 k

1

Temperature coefficient -10

°

C to 18

°

C, +28

°

C to

55

°

C:

±

0.005 % of resistance range per

°

C