724
Users Manual

26

Using Source Mode

In SOURCE mode, the calibrator generates calibrated
signals for testing and calibrating process instruments,
supplies voltages and resistances, and simulates the
electrical output of RTD and thermocouple temperature
sensors.

Sourcing Electrical Parameters

Volts or ohms are sourced and shown in the lower
display.

To select an electrical sourcing function, proceed as
follows:

1. Connect the test leads as shown in Figure 12,

depending on the source function.

2. If necessary, press

M

for SOURCE mode.

3. Press

V

for dc voltage, or

U

for resistance.

4. Enter the desired output value by pressing

X

and

W

keys. Press

Y

and

Z

to select a different digit to

change.

V

mA

MEASURE

SOURCE / MEASURE

30V MAX ALL TERMINALS

4W

3W

V

RTD

TC

COM

COM

LOOP

Red

Black

+

Common

V

zi09f.eps

Figure 12. Electrical Sourcing Connections