Setting 0 % and 100 % Output
Parameters

For output parameters (volts, ohms, TC potentials or RTD
resistances), you must set the 0 % and 100 % points
before you can use the step and ramp functions. Proceed
as follows:

1. If necessary, press

M

for SOURCE mode.

2. Select the TC source function and use the arrow keys

to enter the value. Our example is thermocouple
source using 100

°

C and 300

°

C values for source.

3. Enter

100

°

C and press and hold

J

to store the

value.

4. Enter in 300

°

C and press and hold

G

to store the

value.

You can now use this setting for the following:

Manually stepping an output with 25 % increments.

Jump between the 0 and 100 % span points by
momentarily pushing

J

or

G

.

Stepping and Ramping the Output

Two features are available for adjusting the value of
source functions.

Stepping the output manually with the

H

and

I

keys, or in automatic mode.

Ramping the output.

Stepping and ramping apply to all functions.

Manually Stepping the Output

To manually step the output you can do the following:

Use

H

or

I

to step the output up or down in

25 % steps.

Touch momentarily either

J

to go to 0 %, or

G

to go to 100 %.