34

Testing an Output Device

Use the source functions to test and calibrate actuators,
recording, and indicating devices. Proceed as follows:

1. Connect the test leads to the instrument under test

as shown in Figure 16.

2. Press

l

for dc voltage, or

U

for resistance

(lower display).

3. If necessary, press

M

for SOURCE mode.

TC

RTD

100%

25%

25%

RECALL

MEAS

SOURCE

STORE

SETUP

0%

TEMPERATURE CALIBRATOR

724

Black

Red

0 to 1 V dc

Input

V

LOOP

mA

V

˚

F

˚

C

zi13f.eps

Figure 16. Calibrating a Chart Recorder