beautypg.com

Auto ramping the output, Storing and recalling setups – Atec Fluke-724 User Manual

Page 41

background image

Temperature Calibrator

Storing and Recalling Setups

31

Auto Ramping the Output

Auto ramping gives you the ability to continuously apply a
varying stimulus from the calibrator to a transmitter, while
your hands remain free to test the response of the
transmitter.

When you press

L

, the calibrator produces a

continuously repeating 0 % - 100 % - 0 % ramp in your
choice of three ramp waveforms:

E

0 % - 100 % - 0 % 40-second smooth ramp

P

0 % - 100 % - 0 % 15-second smooth ramp

N

0 % - 100 % - 0 % Stair-step ramp in 25 %

steps, pausing 5 seconds at each step.

To exit ramping, press any button.

Storing and Recalling Setups

You can store up to eight of your settings in a nonvolatile
memory and recall the settings for later use. A low battery
condition or a battery change does not jeopardize the
stored settings. Proceed as follows:

1. After you create a calibrator setup, press

S

. In the

display, the memory locations appear.

2. Press

Y

or

Z

to select locations one through eight. An

underscore appears below the selected memory
location.

3. Press

S

.

Only the stored memory location will be

displayed. The setup is stored.

To recall setups, proceed as follows.

1. Press

Q

. The memory locations appear on the

display.

2. Press

Y

or

Z

to select the appropriate location and

press

Q

.