Other data – Atec Rohde-Schwarz-SMIQ Series User Manual
Page 29
Vector Signal Generator R&S SMIQ – Specifications
29
Other data
Memory for instrument settings
50 storable settings
List mode
Remote control
General data
Memory sequence modes
automatic, single shot, manual or external trigger
Step time
Resolution
50 ms to 60 s
1 ms
Frequency and level values can be stored in a list and set in an extremely short time; permissible level variation: 90 dB
Modes
automatic, single shot, manual or external trigger
Max. number of channels
2000
Dwell time
Resolution
0.5 ms to 1 s
0.1 ms
System
IEC 60625 (IEEE 488)
Command set
SCPI 1993.0
Connector
24-contact Amphenol
IEC/IEEE-bus address
0 to 30
Interface functions
SH1, AH1, T6, L4, SR1, RL1, PP1, DC1, DT1, C0
System
RS 232
Command set
SCPI 1993.0
Connector
9-contact D-SUB
Baud rate
1.2 kbit/s to 115.2 kbit/s
Power supply
100 V to 240 V (AC)
± 10%
, 50 Hz to 400 Hz,
autosetting to AC supply, max. 300 VA
Electromagnetic compatibility
Immunity to RFI
to EN 55011 and EN 61326-2 (EMC directive of EU)
10 V/m
Ambient conditions
Operating temperature range
Storage temperature range
Climatic conditions
0°C to 50°C
6 )
to IEC60068
−40°C to +70°C
95% relative humidity at +25°C/+40°C cyclically,
to IEC 60068
Mechanical resistance
Sinewave vibrations
5 Hz to 150 Hz, max. 2 g at 55 Hz,
max. 0.5 g in range 55 Hz to 150 Hz,
to IEC 60068, IEC 61010 and MIL-T-28800D, class 5
Random
Shock
10 Hz to 300 Hz, acceleration 1.2 g (rms)
40 g shock spectrum, to MIL-STD-810D, MIL-T-28800D,
class 3 and 5
Safety
to EN 61010-1
Dimensions (W x H x D)
435 mm x 192 mm x 460 mm
Weight