beautypg.com

Other data – Atec Rohde-Schwarz-SMIQ Series User Manual

Page 29

background image

Vector Signal Generator R&S SMIQ – Specifications

29

Other data

Memory for instrument settings

50 storable settings

List mode

Remote control

General data

Memory sequence modes

automatic, single shot, manual or external trigger

Step time

Resolution

50 ms to 60 s
1 ms

Frequency and level values can be stored in a list and set in an extremely short time; permissible level variation: 90 dB
Modes

automatic, single shot, manual or external trigger

Max. number of channels

2000

Dwell time

Resolution

0.5 ms to 1 s
0.1 ms

System

IEC 60625 (IEEE 488)

Command set

SCPI 1993.0

Connector

24-contact Amphenol

IEC/IEEE-bus address

0 to 30

Interface functions

SH1, AH1, T6, L4, SR1, RL1, PP1, DC1, DT1, C0

System

RS 232

Command set

SCPI 1993.0

Connector

9-contact D-SUB

Baud rate

1.2 kbit/s to 115.2 kbit/s

Power supply

100 V to 240 V (AC)

± 10%

, 50 Hz to 400 Hz,

autosetting to AC supply, max. 300 VA

Electromagnetic compatibility
Immunity to RFI

to EN 55011 and EN 61326-2 (EMC directive of EU)
10 V/m

Ambient conditions
Operating temperature range
Storage temperature range
Climatic conditions

0°C to 50°C

6 )

to IEC60068

−40°C to +70°C
95% relative humidity at +25°C/+40°C cyclically,
to IEC 60068

Mechanical resistance
Sinewave vibrations

5 Hz to 150 Hz, max. 2 g at 55 Hz,
max. 0.5 g in range 55 Hz to 150 Hz,
to IEC 60068, IEC 61010 and MIL-T-28800D, class 5

Random
Shock

10 Hz to 300 Hz, acceleration 1.2 g (rms)
40 g shock spectrum, to MIL-STD-810D, MIL-T-28800D,
class 3 and 5

Safety

to EN 61010-1

Dimensions (W x H x D)

435 mm x 192 mm x 460 mm

Weight

25 kg when fully equipped