Vector Signal Generator R&S SMIQ – Specifications

11

Memory extension (option R&S SMIQB12)

The data generator memory can be extended to max. 79 Mbit by fitting up to two options R&S SMIQB12.

Digital standards with options R&S SMIQB20 and R&S SMIQB11

Inputs

Trigger input TRIGIN
Trigger frequency
Selectable trigger delay
Selectable trigger suppression
Switching threshold
Input impedance
Required pulse width

for starting the data sequences in the data memory
<10 kHz
0 to 2

16

−1 symbols

0 to 2

26

−1 symbols after trigger

−2.5 V to 2.5 V, selectable, resolution 0.1 V
1 k

Ω to ground, 50 Ω to ground

>50 ns

Outputs

DATA modulation data
BURST GATE, LEV ATT
CW

see data under “Digital Modulation“
control signals for envelope control and level attenuation
control signal for switching off digital modulation

TRIGOUT 1, TRIGOUT 2
TRIGOUT 3
HOP

user-programmable trigger signals
trigger signal on event
control signal for frequency hopping in LIST MODE

Memory capacity

32 Mbit

Modulation and ACP specifications apply at the respective standard frequency ranges and at levels
≤8 dBm PEP with R&S SMIQ02B/03B

≤5 dBm PEP with R&S SMIQ04B/06B

Data sources

with TDMA
with CDMA

internal, external via RS232 interface (SERDATA)
internal

Digital standard GSM / EDGE

Frequency

Standard
Range

to GSM standard

880 MHz to 960 MHz/1710 MHz to 2000 MHz
same as R&S SMIQ

Modulation

GMSK or 8PSK EDGE (8PSK with 3

π/8 rotation)

Symbol rate

Standard
Range

270 833 Hz
100 Hz to 300 kHz

Baseband filter (GMSK)

Standard
Range

Gauss, BxT = 0.3
Gauss, BxT = 0.2 to 0.7

Baseband filter (8PSK EDGE)

Gauss Linear

Frame structure

1 to 8 slots user-definable
any burst types can be combined, incl. GSM and EDGE
mixed

Burst types

Burst on/off ratio
Burst rise/fall time
Slot attenuation, selectable

NORMAL, DUMMY, ALL DATA, NORMAL EDGE
>80 dB
corresponding to 3 symbols,

<18 µs

0 dB to 70 dB

Internal modulation data

PRBS, 2

9

−1, 2

11

−1, 2

15

−1, 2

16

−1, 2

20

−1, 2

21

−1, 2

23

−1,

programmable data memory

External serial modulation data

asynchronous via RS232 interface (SERDATA)