Plz-3w/3wh series, Electronic load (cc·cr·cv·cp), Specifications(plz-3wh series) – Atec Kikusui-PLZ1003W User Manual
Page 7

Copyright © 2001 KIKUSUI ELECTRONICS CORP. All rights reserved.
http://www.kikusui.co.jp/
Page7
PLZ-3W/3WH series
ELECTRONIC LOAD (CC·CR·CV·CP)
Specifications(PLZ-3WH series)
■
Operating Area
PLZ153WH
PLZ303WH
PLZ603WH
PLZ1003WH
Operating voltage
(DC)*1
5 to 500 V
Current
7.5 A
15 A
30 A
50 A
Power
150 W
300 W
600 W
1000 W
Minimum operation-
starting voltage*2
1 V
*1 Current can flow in a range of 1 to 5 V. However, the specifications for this instru
ment may not be met.
*2 Minimum voltage at which current starts to flow in the instrument
■
Constant Voltage Mode
PLZ153WH
PLZ303WH
PLZ603WH
PLZ1003WH
Operating range (DC)
5 V to 500 V
Setting accuracy*1
±
0.1% of rated voltage
Setting resolution
125 mV
DC input current
regulation*2
±
0.01% of rated voltage
Temperature coefficient
±
100 ppm/
°
C of rated voltage (typical value)
*1 In a range of 23
±
5
°
C (with the load terminals remote sensed)
*2 With respect to an input current change of 10% to 100% of the rated current at 5 V
input voltage (with the load terminals remote sensed)
■
Constant Power Mode
PLZ153WH
PLZ303WH
PLZ603WH
PLZ1003WH
Operating range
15 to 150 W
30 to 300 W
60 to 600 W
100 to 1000 W
Setting accuracy*1
±
2% of rated power
Setting resolution
0.025% of rated power
DC input voltage regulation*2
2% of rated power
Ripple current*3
Complies with graph 2 (typical values)
Temperature coefficient
±
1000 ppm/
°
C of rated power (typical value)
*1 In a range of 23
±
5
°
C and at 20 V input voltage (with the load terminals remote sensed)
*2 With respect to an input voltage change of 20 V to 500 V at rated power (with the load
terminals remote sensed)
*3 When the ripple noise of input voltage is 5 mV or less
■
Constant Resistance Mode
PLZ153WH
PLZ303WH
PLZ603WH
PLZ1003WH
Operating range
Range H*5
1.6
Ω
to 20k
Ω
0.8
Ω
to 10k
Ω
0.4
Ω
to 5k
Ω
0.24
Ω
to 3k
Ω
0.625 to
1.25 to
2.5 to
4.17 to
5
×
10
-5
S
1×
10
-4
S
2×
10
-4
S
3.3×
10
-4
S
Range L*5
16
Ω
to 200k
Ω
8
Ω
to 100k
Ω
4
Ω
to 50k
Ω
2.4
Ω
to 30k
Ω
0.0625 to
0.125 to
0.25 to
0.417 to
5×
10
-6
S
1×
10
-5
S
2×
10
-5
S
3.3×
10
-5
S
Setting resolution
Range H*6
0.156 mS
0.3125 mS
0.625 mS
1 mS
Range L*6
0.0156 mS
0.03125 mS
0.0625 mS
0.1 mS
Setting accuracy
(current conversion)*1
±
(1% of rated current +a)*3
Stability
DC input voltage regulation*2
6%
Temperature
±
(1000 ppm/
°
C + 4 m
Ω
/
°
C)
coefficient
at minimum resistance value
Ripple noise*4
Complies with graph 1 (typical values)
*1 In a range of 23
±
5
°
C and at 12 V input voltage
*2 With respect to an input voltage change of 5 V to 12 V at the minimum resistance
value. For other resistance, the voltage variations will be within 6% of the maximum
conductance (S) value of the setting range with respect to variations in allinput
voltages. (with the load terminals remote sensed)
*3
α
= 12 V
÷
set resistance - 12 V
÷
(set resistance + 120 m
Ω
)
*4 A repetitive noise of about 155 kHz may be superimposed on input current.
*5 S:siemens
*6 mS: millisiemens
Note: Conductance (S) x input voltage (V) = load current (A)
Conductance (S) = 1/resistance (
Ω
)
■
Constant Current Mode
PLZ153WH
PLZ303WH
PLZ603WH
PLZ1003WH
Operating range
Range H
0 to 7.5 A
0 to 15 A
0 to 30 A
0 to 50 A
Range L
0 to 0.75 A
0 to 1.5 A
0 to 3 A
0 to 5 A
Setting accuracy
(with respect to rated
±
(0.3%
±
(0.3%
±
(0.3%
±
(0.3%
current)*1
+ 7.5 mA)
+ 15 mA)
+ 30 mA)
+ 50 mA)
Setting resolution
Range H
2 mA
4 mA
8 mA
13 mA
Range L
0.2 mA
0.4 mA
0.8 mA
1.3 mA
Stability
Line regulation*2
3 mA
DC input voltage regulation*3
20 mA
Temperature coefficient
±
100 ppm/
°
C of rated current (typical value)
Ripple noise*4
RMS*5
2 mA
2 mA
3 mA
5 mA
Measuring current*6
At 7.5 A
At 15 A
At 30 A
At 50 A
Peak-to-peak
20 mA
20 mA
30 mA
50 mA
Measuring current
At 7.5 A
At 15 A
At 30 A
At 50 A
*1 In a range of 23
±
5
°
C
*2 With respect to a variation of
±
10% of the center value of the input supply voltage at
the rated current of 20 V input voltage
*3 Value obtained when input voltage is varied from 5 V to 500 V at the current of rated
power/500 V
*4 At the rated current of 20 V input voltage
*5 5 Hz to 500 kHz
*6 DC to 15 MHz
■
Tr/Tf, Switching and Soft-start Operations, Remote Sensing, and Protective Features
PLZ153WH
PLZ303WH
PLZ603WH
PLZ1003WH
Tr/Tf setting*1
Operation mode
Constant current
Setting range
50, 100, 200, 500
µ
s
1, 2, 5, 10 ms
Setting accuracy*2
±
30% of set value,
±
15
µ
s
Switching operation
Operation mode
Constant current, constant resistance
Time setting range
1 to 5000 ms
Time setting accuracy
±
5% of set value
Soft-start operation
Operation mode
Constant current(C.C)
Setting range
0.5, 1, 2, 5, 10, 20, 50, 100 ms
Setting accuracy
±
30% of set value,
±
100
µ
s
Remote sensing
Sensing voltage
5 V*3
Protective features
Over current protection
Limit is activated at about
(OCP)
+5% of rated current.
Overheat protection
Load current is cut off at a he
(OHP)
at sink temperature of about 105
°
C.
Reverse connection
protection
With diodes and fuses
Power transistor
protection
With fuses
Overvoltage protection
(OVP)
Load switch will be turned off.
*1 Within an input voltage range of 12 V to 500 V.
Rise time (Tr) and fall time (Tf) are times required to reach 10% to 90% of current
waveform.
*2 Tr/Tf setting is valid when the changes in load current fall within a range of 2% to
100% of the rated current value.
*3 2.5 V at one terminal