Pcr-l series, Multi purpose ac power supply (cv·cf), Functions

Page 7

background image

Page7

Copyright © 2002 KIKUSUI ELECTRONICS CORP. All rights reserved.

http://www.kikusui.co.jp/

4

Sequence Operation

or

/

Setting the times for the sequence pattern of voltage, frequency or
impedance allows that sequence pattern to be run, stopped or repeated
as desired.

7

Output Impedance Setting

or

/

The output impedance of the PCR-L series is nearly 0

. However,

actual commercial power supplies have an impedance (resistance)
on the order of several m

. The members of the PCR-L series there-

fore allow output impedance to be varied to closely match actual
power supply environments.

8

Power Factor, VA and
Peak Hold Current Measurements

or

/

In addition to these power supplies being equipped with a diverse
range of measuring functions, those functions can be further expanded
by connecting various options. Expanded functions consist of "power
factor measurement", "VA measurement" and "peak hold current
measurement". Measurement of peak hold current involves measur-
ing the peak current until the power supply receives a peak clear
signal or command. This is useful for measuring rush current and so
on.

9

AC+DC Mode

or

/

This function allows output of voltage waveforms in which AC volt-
age is superimposed on DC voltage. This mode allows these power
supplies to be used not only for various electronic devices, but for
chemical experiments and production equipment as well.

10

Memory Function Expansion

or

/

These power supplies allows storage of 9 sets of voltage and fre-
quency settings as a standard feature (at memory addresses 1-9). These
settings can then be read out and output as necessary. The adding of
a memory expansion option allows up 99 sets of voltage and fre-
quency settings to be stored in memory. In the AC+DC mode, AC
memory values and DC memory values are output simultaneously
thereby allowing memory operation in the AC+DC mode.

11

Regulation Adjustment

This function is used for the same purpose as the "sensing function".
The sensing function measures the voltage at a specific sensing point
and then holds the voltage of that sensing point constant. In contrast,
the regulation adjustment function calculates the voltage drop caused
by output current and then raises the output voltage by the amount of
that drop. This function is used for stabilizing the voltage at the load
end when the power supply and load are located at a considerable
distance from each other.

Note: When using the regulation adjustment function, the voltage stabilization accuracy,

distortion factor and response speed will decrease below the level of performance
of the power supply.

100 V Range

Variable Range

Resolution

Error

PCR500L

0 to 4.0

40m

±

(20%+80m

)

PCR1000L

0 to 2.0

20m

±

(20%+40m

)

PCR2000L

0 to 1.0

10m

±

(20%+20m

)

PCR4000L

0 to 0.5

5m

±

(20%+10m

)

PCR6000L

0 to 0.333

3.33m

±

(20%+6.67m

)

200V Range

PCR500L

0 to 16.0

160m

±

(20%+320m

)

PCR1000L

0 to 8.0

80m

±

(20%+160m

)

PCR2000L

0 to 4.0

40m

±

(20%+80m

)

PCR4000L

0 to 2.0

20m

±

(20%+40m

)

PCR6000L

0 to 1.333

13.33m

±

(20%+26.67m

)

Functions

5

Output Phase Setting

or

/

The timing by which the output is
switched on and off can be arbi-
trarily set with the phase of the volt-
age waveform. Combining this
phase setting function with the cur-
rent peak hold function greatly fa-
cilitates measurement of rush cur-
rent.

Output ON/OFF Phase Setting
Output on phase: 90 deg.
Output off phase: 0 deg

6

Harmonics Current Analysis Function

or

/

This function performs analysis of harmonics current for output cur-
rent from the power supply. Although this measurement was con-
ventionally measured using an FFT analyzer, the use of these power
supplies eliminates the need for this device. In addition, the use of a
standardized line impedance network as shown in the diagram below
also allows simulation of commercial power supplies.

System Concept Drawing

For further information, contact KIKUSUI distributors.

Effective Mode

AC

DC

AC+DC

Address: ADRS

0 to 99

:HOUR

0 to 999 h 59 min.

Time

:MIN

0 to 999 min. 59 s

:SEC

0 to 999.999s

Waveform:WAVE

Sine, peak clip

Voltage (AC): Vac

1.0V to 300.0 VAC

Frequency: FRQ

1.00Hz to 999.9 Hz

Impedance (resistance):

Varies according

IMP

to model

Frequency change

Ramp/step

characteristics

Voltage (DC): Vdc

-424.0V to + 424.0V

Status signal: STAT

ON/OFF

Trigger signal: TRG

ON/OFF

Output: OUT

ON/OFF

AC voltage change

Ramp/step

characteristics

Setteing

Setting and Range

(allows input of
14 waveforms)

Printer

GPIB

RS-232C

PCR-L Series

Computer

AC Power Supply

FFT Analysis Function
Power factor/electrical power
measurement function
Current/voltage measurement
function

Line

Impedance

Network

Tested

Equipment

Software

"PCR-L Applications with Linzo" (SD02-PCR-L)

PCR-L series

MULTI PURPOSE AC POWER SUPPLY (CV·CF)