Pp64 bandwidth, Error power, Error of amplitude – Atec Fluke-Norma-4000-5000 Series User Manual
Page 9
9
ENEN
Fluke Corporation Fluke Norma 4000/5000 power analyzers
Voltage and current depending on the input
level at 45 – 65 Hz frequency range
Input level
Sum limit of error U
Sum limit of error I
in %
in %
in %
100
0.03
0.03
50
0.05
0.05
30
0.08
0.08
10
0.21
0.21
5
0.41
0.41
3
0.68
0.68
1
2.01
0.03
PP64 Bandwidth
U and I via BNC
I direct measured
Bandwidth –3dB
3 MHz
1 MHz
Frequency
Limits of error current [ % ]
[Hz]
Internal shunt
[Hz]
0
0.03
-0.03
0.03
-0.03
10
0.03
-0.03
0.03
-0.03
45
0.03
-0.03
0.03
-0.03
65
0.03
-0.03
0.03
-0.03
500
0.03
-0.03
0.03
-0.03
1000
0.03
-0.03
0.03
-0.03
2000
0.08
-0.21
0.06
-0.15
5000
0.13
-0.45
0.10
-0.31
10000
0.18
-0.63
0.13
-0.43
20000
0.23
-0.81
0.16
-0.55
50000
0.28
-1.05
0.20
-0.71
100000
0.33
-1.23
0.23
-0.83
Current measurement accuracy depending on
the frequency (reading + range)
Power / Angular Error
PP64
Between U and I
BNC
Between U and I
Shunt
Angular Error
0.0025° +0.0025°/kHz
0.005° +0.005°/kHz
Frequency in Hz
Limit of error power in % at Lamda = 1
45
0.05
-0.05
65
0.05
-0.05
Error Power
Current
6 ranges:
30 – 100 mA – 0,3 – 1 – 3 – 10 A
I
peak
= 2 x range; max level 150% at
sine wave (limit of error as at 100%).
Input impedance with integrated shunts:
Ranges
30, 100 mA:
1 Ohm
Ranges
0.3, 1 A:
0.1 Ohm
Ranges
3, 10 A:
0.01 Ohm
Current overload:
max. 15 A continuous
30 A < 5 sec / 15 sec no load
100 A < 0.1 s / 30 sec no load
Input for external shunt or probe:
BNC terminal:
100 kOhm / 30 pF
30 – 100 mV – 0.3 – 1 – 3- 10 V
Overload:
max. 20 V
rms
CMR common mode rejection:
120 dB at 100 kHz
Error of Amplitude
Basic accuracy PP64
Sum limit of error
U
I
range
0.02%
0.02%
reading
0.01%
0.01%
Frequency in Hz
Limits of error voltage in %
0
0.03
-0.03
10
0.03
-0.03
45
0.03
-0.03
65
0.03
-0.03
500
0.03
-0.03
1000
0.06
-0.15
5000
0.10
-0.31
10000
0.13
-0.43
20000
0.16
-0.55
50000
0.20
-0.71
100000
0.23
-0.83
Valid for averaged values at 23 ± 0.5 °C ambient temperature,
sine waveform and after 1 hour turn on time with measuring
signal.
Voltage measurement accuracy
depending on the frequency (reading + range)