Pp64 bandwidth, Error power, Error of amplitude – Atec Fluke-Norma-4000-5000 Series User Manual

Page 9

background image

9

ENEN

Fluke Corporation Fluke Norma 4000/5000 power analyzers

Voltage and current depending on the input

level at 45 – 65 Hz frequency range

Input level

Sum limit of error U

Sum limit of error I

in %

in %

in %

100

0.03

0.03

50

0.05

0.05

30

0.08

0.08

10

0.21

0.21

5

0.41

0.41

3

0.68

0.68

1

2.01

0.03

PP64 Bandwidth

U and I via BNC

I direct measured

Bandwidth –3dB

3 MHz

1 MHz

Frequency

Limits of error current [ % ]

[Hz]

Internal shunt

[Hz]

0

0.03

-0.03

0.03

-0.03

10

0.03

-0.03

0.03

-0.03

45

0.03

-0.03

0.03

-0.03

65

0.03

-0.03

0.03

-0.03

500

0.03

-0.03

0.03

-0.03

1000

0.03

-0.03

0.03

-0.03

2000

0.08

-0.21

0.06

-0.15

5000

0.13

-0.45

0.10

-0.31

10000

0.18

-0.63

0.13

-0.43

20000

0.23

-0.81

0.16

-0.55

50000

0.28

-1.05

0.20

-0.71

100000

0.33

-1.23

0.23

-0.83

Current measurement accuracy depending on

the frequency (reading + range)

Power / Angular Error

PP64

Between U and I

BNC

Between U and I

Shunt

Angular Error

0.0025° +0.0025°/kHz

0.005° +0.005°/kHz

Frequency in Hz

Limit of error power in % at Lamda = 1

45

0.05

-0.05

65

0.05

-0.05

Error Power

Current
6 ranges:

30 – 100 mA – 0,3 – 1 – 3 – 10 A

I

peak

= 2 x range; max level 150% at

sine wave (limit of error as at 100%).

Input impedance with integrated shunts:
Ranges

30, 100 mA:

1 Ohm

Ranges

0.3, 1 A:

0.1 Ohm

Ranges

3, 10 A:

0.01 Ohm

Current overload:

max. 15 A continuous
30 A < 5 sec / 15 sec no load
100 A < 0.1 s / 30 sec no load

Input for external shunt or probe:
BNC terminal:

100 kOhm / 30 pF

30 – 100 mV – 0.3 – 1 – 3- 10 V

Overload:

max. 20 V

rms

CMR common mode rejection:

120 dB at 100 kHz

Error of Amplitude

Basic accuracy PP64
Sum limit of error

U

I

range

0.02%

0.02%

reading

0.01%

0.01%

Frequency in Hz

Limits of error voltage in %

0

0.03

-0.03

10

0.03

-0.03

45

0.03

-0.03

65

0.03

-0.03

500

0.03

-0.03

1000

0.06

-0.15

5000

0.10

-0.31

10000

0.13

-0.43

20000

0.16

-0.55

50000

0.20

-0.71

100000

0.23

-0.83

Valid for averaged values at 23 ± 0.5 °C ambient temperature,

sine waveform and after 1 hour turn on time with measuring

signal.

Voltage measurement accuracy

depending on the frequency (reading + range)