Pp54 power phase ranges, Error of amplitude, Pp54 bandwidth – Atec Fluke-Norma-4000-5000 Series User Manual
Page 8: Power / angular error, Error power, Pp64 power phase ranges
8
Voltage
8 ranges:
0.3 – 1 – 3 – 10 – 30 – 100 – 300 –
1000 V
U
peak
= 2 x range
Input impedance:
2 MOhm / 20 pF
CMR common mode rejection:
120 dB at 100 kHz
PP54 Power Phase
Ranges
ENEN
Fluke Corporation Fluke Norma 4000/5000 power analyzers
Current
6 ranges:
30 – 100 mA – 0,3 – 1 – 3 – 10 A
I
peak
= 2 x range; max level 150% at
sine wave (limit of error as at 100%).
Input impedance with integrated shunts:
Ranges
30, 100 mA:
1 Ohm
Ranges
0.3, 1 A:
0.1 Ohm
Ranges
3, 10 A:
0.01 Ohm
Current overload:
max. 15 A continuous
30 A < 5 sec / 15 sec no load
100 A < 0.1 s / 30 sec no load
Input for external shunt or probe:
BNC terminal:
100 kOhm / 30 pF
30 – 100 mV – 0.3 – 1 – 3- 10 V
Overload:
max. 20 V
rms
CMR common mode rejection:
120 dB at 100 kHz
Error of Amplitude
Basic accuracy PP54
Sum limit of error
U
I
range
0.05%
0.05%
reading
0.05%
0.05%
Voltage and current depending on the input
level at 45 – 65 Hz frequency range
Input level
Sum limit of error U
Sum limit of error I
in %
in %
in %
100
0.10
0.10
50
0.15
0.15
30
0.22
0.22
10
0.55
0.55
5
1.05
1.05
3
1.72
1.72
1
5.05
5.05
PP54 Bandwidth
U and I via BNC
I direct measured
Bandwidth –3 dB
3 MHz
1 MHz
Current measurement accuracy
depending on the frequency (reading + range)
Frequency
Limits of error current [ % ]
[Hz]
Internal shunt
[Hz]
0
0.10
-0.10
0.10
-0.10
10
0.10
-0.10
0.10
-0.10
45
0.10
-0.10
0.10
-0.10
65
0.10
-0.10
0.10
-0.10
500
0.10
-0.10
0.10
-0.10
1000
0.10
-0.10
0.10
-0.10
2000
0.15
-0.28
0.13
-0.22
5000
0.20
-0.52
0.17
-0.38
10000
0.25
-0.70
0.20
-0.50
20000
0.30
-0.88
0.23
-0.62
50000
0.35
-1.12
0.27
-0.78
100000
0.40
-1.30
0.30
-0.90
Voltage measurement accuracy depending on
the frequency (reading + range)
Frequency in Hz
Limits of error voltage in %
0
0.10
-0.10
10
0.10
-0.10
45
0.10
-0.10
65
0.10
-0.10
500
0.10
-0.10
1000
0.10
-0.10
2000
0.13
-0.22
5000
0.17
-0.38
10000
0.20
-0.50
20000
0.23
-0.62
50000
0.27
-0.78
100000
0.30
-0.90
Valid for averaged values at 23 ± 0.5 °C ambient
temperature, sine waveform and after 1 hour turn on time
with measuring signal.
Power / Angular Error
PP54
Between U and I
BNC
Between U and I
Shunt
Angular Error
0.005° +0.005° / kHz
0.025° + 0.015° / kHz
Aliasing filter OFF
Error Power
Frequency in Hz
Limit of error power in % at Lamda = 1
45
0.16
-0.16
65
0.16
-0.16
Voltage
8 ranges:
0.3 – 1 – 3 – 10 – 30 – 100 – 300 –
1000 V
U
peak
= 2 x range
Input impedance:
2 MOhm / 20 pF
CMR common mode rejection:
120 dB at 100 kHz
PP64 Power Phase
Ranges