Specifications – Atec Fluke-Norma-4000-5000 Series User Manual
Page 5
ENEN
Ambient conditions
Working Temperature Range
5 °C to 35 °C (41 °F to 95 °F)
Storage Temperature Range
-20 °C to 50 °C (-4 °F to 122 °F)
Housing
Equipped with a solid metal case to meet stringent EMC requirements.
Weight
Fluke Norma 4000: Base unit approx. 5 kg (11 lb.)
Fluke Norma 5000: Base unit approx. 7 kg (15 lb.)
Size (HxWxD)
Fluke Norma 4000:
15 cm x 23.7 cm x 31.5 cm (5.9 in. x 9.3 in. x 12.4 in.)
Fluke Norma 5000:
15 cm x 44.7 cm x 31.5 cm (5.9 in. x 17.6 in x 12.4 in.)
Display
5.7“ / 144mm - 320 x 240 pixel
User-selectable background lighting and contrast.
Climatic Class
KYG DIN 40040, max. 85 % relative humidity,
non-condensing.
Net Connection
85 to 264 V ac, 50 to 60 Hz, dc 100 to 260 V, ca. 40 VA European plug with switch.
Measuring Inputs
Safety sockets 4 mm, 2 for each input. External shunt connection over BNC socket.
Operation
Membrane keyboard with cursor – function keys and direct functions.
Connections
Rear panel of the 3-phase analyzer
Measured Values
Non-gapping calculation of averaged values for each phase. In three phase system additionally calculation of total
power and averaging of V and I of the three phases. The fundamental H01 will be calculated in synchronous mode
also for these values.
U
RMS
effective value, U
rm
rectified mean, U
m
mean value
U
p
-, U
p
+, U
pp
peak values
U
cf
crest factor U
cf
, U
ff
form factor
U
fc
fundamental content
U
thd
distortion factor DIN, IEC
I
RMS
effective value, I
rm
rectified mean, I
m
mean value
I
p-
, I
p+
, I
pp
peak values
I
cf
crest factor I
cf
, I
ff
form factor
I
fc
fundamental content
I
thd
distortion factor DIN, IEC
P active power [W]
Q reactive power [Var]
S apparent power [VA]
ë, cos. phase angular
Integral function for active power P, reactive power Q, apparent power S, voltage (U
m
) and current (I
m
),
Number of digits 4 or 5 dependent on measured value.
Frequency and Synchronization
Range
DC and 0.2 Hz to sampling rate
Accuracy
±0.01 % of measured value (reading)
•
Channels which can be selected: all U/I or external input.
•
One of three low pass filter with different frequencies can be switched into the signal.
•
The frequency is always visible on the top of the screen.
•
The BNC synchronization socket on backside of the instrument can be used either as input or output.
•
The input signals can be measured up to the sample rate of the power phase. The maximum level must not be higher then 50 V.
•
The output signal is a pulsed 5 Volts TTL signal (frequency depends on the measured synch frequency).
Configuration Memory
Up to 15 user configurations can be saved into a permanent memory and reloaded later on. Changes that were not saved are lost after switching off the instrument.
Interface
RS232/USB interface for upload of firmware and data exchange with the PC. A printer can be connected over an external converter.
Options
IEEE 488.2 / 1 MBit/s
Ethernet / 10 MBit/s or 100 Mbit/s
Standards and Safety
Electrical Safety
EN 61010-1 / 2
nd
Edition 1000 V CAT II (600 V CAT III)
Degree of pollution 2, safety Class I
EN 61558 for transformer
EN 61010-2-031/032 for accessories
Maximum Inputs
For voltage inputs Measurement range 1000 V
eff
, 2 kV
peak
For current inputs Measurement range 10 A
eff
, 20 A
peak
Test Voltages
Net input - case (protective conductor): 1.5 KV ac
Net connection – Measurement input: 5.4 kV ac
Measurement inputs – Case: 3.3 kV ac
Measurement input – Measurement input: 5.4 kV
Electromagnetic Susceptibility
Emission: IEC 61326-1, EN 50081-1, EN 55011 Class B
Immunity: IEC 61326-1 / Annex A (industrial sector), EN 50082-1
5
Specifications
Fluke Corporation Fluke Norma 4000/5000 power analyzers