Error power, Pp50 power phase ranges, Error of amplitude – Atec Fluke-Norma-4000-5000 Series User Manual

Page 7: Pp50 bandwidth, Power / angular error

background image

ENEN

7 Fluke Corporation Fluke Norma 4000/5000 power analyzers

Error Power

Frequency in Hz

Limit of error power in % at Lamda = 1

45

0.33

-0.33

65

0.33

-0.33

PP50 Power Phase

Ranges

Voltage
8 ranges:

0.3 – 1 – 3 – 10 – 30 – 100 – 300 –

1000 V

U

peak

= 2 x range

Input impedance:

2 MOhm / 20 pF

CMR common mode rejection:

120 dB at 100 kHz

Current
6 ranges:

30 – 100 mA – 0.3 – 1 – 3 – 10 A

I

peak

= 2 x range; max level 150% at

sine wave (limit of error as at 100%).

Input impedance with integrated shunts:
Ranges 30, 100 mA:

1 Ohm

Ranges 0.3, 1 A:

0.1 Ohm

Ranges 3, 10 A:

0.01 Ohm

Current overload:

max. 15 A continuous
30 A < 5 sec / 15 sec no load
100 A < 0.1 s / 30 sec no load

Input for external shunt or probe:
BNC terminal:

100 kOhm / 30 pF

30 – 100 mV – 0.3 – 1 – 3- 10 V

Overload:

max. 20 Vrms

CMR common mode rejection:

120 dB at 100 kHz

Error of Amplitude

Basic accuracy PP50
Sum limit of error

U

I

range

0.05%

0.05%

reading

0.05%

0.05%

Voltage and current depending on the input

level at 45 – 65 Hz frequency range

Input level

Sum limit of error U

Sum limit of error I

in %

in %

in %

100

0.10

0.10

50

0.15

0.15

30

0.22

0.22

10

0.55

0.55

5

1.05

1.05

3

1.72

1.72

1

5.05

5.05

PP50 Bandwidth

U and I via BNC

I direct measured

Bandwidth –3 dB

10 MHz

1 MHz

Current measurement accuracy

depending on the frequency (reading + range)

Frequency

Limits of error current [ % ]

[Hz]

Internal shunt

BNC input

0

0.10

-0.10

0.10

-0.10

10

0.10

-0.10

0.10

-0.10

45

0.10

-0.10

0.10

-0.10

65

0.10

-0.10

0.10

-0.10

500

0.10

-0.10

0.10

-0.10

1000

0.10

-0.10

0.10

-0.10

2000

0.12

-0.19

0.12

-0.16

5000

0.15

-0.31

0.13

-0.24

10000

0.18

-0.40

0.15

-0.30

20000

0.20

-0.49

0.17

-0.36

50000

0.23

-0.61

0.18

-0.44

100000

0.25

-0.70

0.20

-0.50

Voltage measurement accuracy depending on

the frequency (reading + range)

Frequency in Hz

Limits of error voltage in %

0

0.10

-0.10

10

0.10

-0.10

45

0.10

-0.10

65

0.10

-0.10

500

0.10

-0.10

1000

0.10

-0.10

2000

0.12

-0.16

5000

0.13

-0.24

10000

0.15

-0.30

20000

0.17

-0.36

50000

0.18

-0.44

100000

0.20

-0.50

Valid for averaged values at 23 ± 0.5 °C ambient

temperature, sine waveform and after 1 hour turn on time

with measuring signal.

Power / Angular Error

PP50

Between U and I

BNC

Between U and I

Shunt

Angular Error

0.005° +0.005° / kHz

0.025° +0.015° / kHz

Aliasing filter OFF

Error Power

Frequency in Hz

Limit of error power in % at Lamda = 1

45

0.16

-0.16

65

0.16

-0.16

Voltage measurement accuracy depending on

the frequency (reading + range)

Frequency in Hz

Limits of error voltage in %

0

0.20

-0.20

10

0.20

-0.20

45

0.20

-0.20

65

0.20

-0.20

500

0.20

-0.20

1000

0.20

-0.20

2000

0.23

-0.32

5000

0.27

-0.48

10000

0.30

-0.60

20000

0.33

-0.72

50000

0.37

-0.88

100000

0.40

-1.00

Valid for averaged values at 23 ± 0,5 °C ambient temperature,

sine waveform and after 1 hour turn on time with measuring

signal.

Power / Angular Error

PP42

Between U and I

BNC

Between U and I

Shunt

Angular Error

0.005° +0.005° / kHz

0.025° +0.015° / kHz

Aliasing filter OFF