Error power, Pp50 power phase ranges, Error of amplitude – Atec Fluke-Norma-4000-5000 Series User Manual
Page 7: Pp50 bandwidth, Power / angular error
ENEN
7 Fluke Corporation Fluke Norma 4000/5000 power analyzers
Error Power
Frequency in Hz
Limit of error power in % at Lamda = 1
45
0.33
-0.33
65
0.33
-0.33
PP50 Power Phase
Ranges
Voltage
8 ranges:
0.3 – 1 – 3 – 10 – 30 – 100 – 300 –
1000 V
U
peak
= 2 x range
Input impedance:
2 MOhm / 20 pF
CMR common mode rejection:
120 dB at 100 kHz
Current
6 ranges:
30 – 100 mA – 0.3 – 1 – 3 – 10 A
I
peak
= 2 x range; max level 150% at
sine wave (limit of error as at 100%).
Input impedance with integrated shunts:
Ranges 30, 100 mA:
1 Ohm
Ranges 0.3, 1 A:
0.1 Ohm
Ranges 3, 10 A:
0.01 Ohm
Current overload:
max. 15 A continuous
30 A < 5 sec / 15 sec no load
100 A < 0.1 s / 30 sec no load
Input for external shunt or probe:
BNC terminal:
100 kOhm / 30 pF
30 – 100 mV – 0.3 – 1 – 3- 10 V
Overload:
max. 20 Vrms
CMR common mode rejection:
120 dB at 100 kHz
Error of Amplitude
Basic accuracy PP50
Sum limit of error
U
I
range
0.05%
0.05%
reading
0.05%
0.05%
Voltage and current depending on the input
level at 45 – 65 Hz frequency range
Input level
Sum limit of error U
Sum limit of error I
in %
in %
in %
100
0.10
0.10
50
0.15
0.15
30
0.22
0.22
10
0.55
0.55
5
1.05
1.05
3
1.72
1.72
1
5.05
5.05
PP50 Bandwidth
U and I via BNC
I direct measured
Bandwidth –3 dB
10 MHz
1 MHz
Current measurement accuracy
depending on the frequency (reading + range)
Frequency
Limits of error current [ % ]
[Hz]
Internal shunt
BNC input
0
0.10
-0.10
0.10
-0.10
10
0.10
-0.10
0.10
-0.10
45
0.10
-0.10
0.10
-0.10
65
0.10
-0.10
0.10
-0.10
500
0.10
-0.10
0.10
-0.10
1000
0.10
-0.10
0.10
-0.10
2000
0.12
-0.19
0.12
-0.16
5000
0.15
-0.31
0.13
-0.24
10000
0.18
-0.40
0.15
-0.30
20000
0.20
-0.49
0.17
-0.36
50000
0.23
-0.61
0.18
-0.44
100000
0.25
-0.70
0.20
-0.50
Voltage measurement accuracy depending on
the frequency (reading + range)
Frequency in Hz
Limits of error voltage in %
0
0.10
-0.10
10
0.10
-0.10
45
0.10
-0.10
65
0.10
-0.10
500
0.10
-0.10
1000
0.10
-0.10
2000
0.12
-0.16
5000
0.13
-0.24
10000
0.15
-0.30
20000
0.17
-0.36
50000
0.18
-0.44
100000
0.20
-0.50
Valid for averaged values at 23 ± 0.5 °C ambient
temperature, sine waveform and after 1 hour turn on time
with measuring signal.
Power / Angular Error
PP50
Between U and I
BNC
Between U and I
Shunt
Angular Error
0.005° +0.005° / kHz
0.025° +0.015° / kHz
Aliasing filter OFF
Error Power
Frequency in Hz
Limit of error power in % at Lamda = 1
45
0.16
-0.16
65
0.16
-0.16
Voltage measurement accuracy depending on
the frequency (reading + range)
Frequency in Hz
Limits of error voltage in %
0
0.20
-0.20
10
0.20
-0.20
45
0.20
-0.20
65
0.20
-0.20
500
0.20
-0.20
1000
0.20
-0.20
2000
0.23
-0.32
5000
0.27
-0.48
10000
0.30
-0.60
20000
0.33
-0.72
50000
0.37
-0.88
100000
0.40
-1.00
Valid for averaged values at 23 ± 0,5 °C ambient temperature,
sine waveform and after 1 hour turn on time with measuring
signal.
Power / Angular Error
PP42
Between U and I
BNC
Between U and I
Shunt
Angular Error
0.005° +0.005° / kHz
0.025° +0.015° / kHz
Aliasing filter OFF